Cleaning & Maintenance
Please follow this procedure to clean the sensor:
Clean any excess oil from the end of the sensor with absorbent paper
Remove the remaining oil by spraying Loctite 7063 cleaner (other low residue cleaners
may also be suitable, please refer to your dealer for more information):
Into each of the four holes at the end of the sensor.
Into the end of the sensor.
All over the outside of the tip.
Remove any excess cleaner, ideally with an airline or with a quick shake of the sensor.
Blow-dry the end of the sensor using a hot-air gun (set to a low temperature (60 to 80°C)
for one minute to ensure no solvent or moisture remains. Alternatively leave sensor to air
dry for 5 minutes.
NOTE: The sensor does not require cleaning once it has been installed unless the application
is particularly high in particulate contamination.
7.8
Physical Dimensions
Notes:
Overall cable length: 2m (nominal)
Strip & tin:5mm
Strip back sheath: 50mm
The M6 lug must be connected by 4sq.mm minimum earth cable to the equipotential
bonding system
Figure 21 - EX Sensor Physical Dimensions