background image

Installing the Battery

Recording will start after the battery is inserted.

Explanation of Symbols

<Warning Symbols>

DANGER

These entries are actions that, if taken, may cause serious personal 

physical damage or death.

CAUTION

These entries are actions that if taken may lead to physical injury or 

damage to persons or things.

<Picture Symbols>

Denotes an important 

warning or caution.

Denotes a 

forbidden action.

Denotes an action 

that should be 

carried out.

 DANGER

Do not take apart, repair or modify the Unit. 

It may result in malfunction or unexpected accidents.

Do not use any batteries other than those specified in this Introductory 

Manual. 

Doing so may cause fire or malfunction.

If water or a foreign object enters into the Unit, immediately remove the battery 

and stop using.

It may result in malfunction or unexpected accidents.

Store the Unit and accessories out of the reach of children. Touching them 

may result in unexpected accidents.

Touching them may result in unexpected accidents.

If any smoke, strange smells or sounds are emitted from the Unit, immediately 

remove the battery and stop using.

Continued use may cause fire or electrocution.

Please be careful not to touch the Unit during or after use in overly hot or cold 

environments; it may cause burns or frostbite.

 CAUTION

We are not responsible for any damage, malfunction or trouble, whether direct 

or indirect, caused by the use of the Unit.Please be fully aware of this before 

using our product.
This Unit has been designed for private and/or industrial use only.It is not for 

use in situations where strict precautions are necessary such as in 

connection with medical equipment, where directly or indirectly.
Battery life varies depending upon measuring environment, frequency of 

communication, Unit settings, and battery performance.
The TR-52i becomes waterproof (splash proof) only after the temperature 

sensor has been connected.

Without the sensor connected, neither the sensor jack of the Unit nor the connector part of the 
temperature sensor is waterproof; make sure not to get wet.

Contact with oil may cause cracks to appear in the casing of the Unit.When 

using the Unit in such an environment, protect the Unit by placing it inside a 

polyethylene bag.

Do not use or store the Unit in such places as listed below; it may cause 
electrocution, fire or damage to the Unit or to your computer.

-Areas exposed to direct sunlight

-Areas exposed in water or high-pressure water flow

-Areas exposed to organic solvents and corrosive gas

-Areas exposed to strong magnetic fields

-Areas exposed to static electricity.

-Areas near fire or exposed to excessive heat

-Areas exposed to excessive dust and smoke

Do not put fingers or foreign objects into the sensor jack.

Do not drop or expose the Unit to a strong impact.

It may cause damage.

    Notices about this Introductory Manual

In order to properly use this product, please carefully read this manual before using.

T&D Corporation accepts no responsibility for any malfunction of and/or trouble with this product or with your computer that is caused by the improper handling of this product and will deem such trouble or 
malfunction as falling outside the conditions for free repair outlined in the attached warranty.
  -   All rights of this Introductory Manual belong to T&D Corporation.It is prohibited to use, duplicate and/or arrange a part or whole of the manual without the permission of T&D Corporation.
  -   All registered trademarks, company names, product names and logos mentioned herein or for products being used are the property of T&D Corporation or of their respective owners. TANDD , 

"T&D", and the logo of T&D Corporation are all registered property of T&D Corporation.

  -   Specifications, design and other contents outlined in the manual are subject to change without notice due to continual improvements.
  -   Please follow the safety precautions outlined in the manual carefully. We cannot guarantee nor are we responsible for safety if this product is used in any manner other than was intended.
  -   On-screen messages in this manual may vary slightly from the actual messages.
  -   Please notify the distributor from which you purchased this product or T&D Corporation of any mistakes, errors or unclear explanations in this manual.T&D Corporation accepts no responsibility for any 

damage or loss of income caused by the use of our product.

  -   This product has been designed for private or industrial use only.It is not for use in situations where strict safety precautions are necessary such as in connection with medical equipment, whether directly or indirectly.
  -   We are not responsible for any malfunction or trouble caused by the use of our product or for any problem caused by the use of measurement results of our product. Please be fully aware of this before using our product.
  -   This Introductory Manual cannot be reissued, so please keep it in a safe place.
  -   Please read the warranty and provisions for free repair carefully.

Safety Precautions and Instructions

The following items should be strictly obeyed for the safe usage of this product, and for protecting yourself and other people from bodily harm and/or damage to property.       

To ensure the proper use of this product, we ask that before using it you carefully read, understand and follow the safety rules and precautions as outlined below.

1. 

Remove the screws and open the cover.

Make sure to use the proper size and type of 
screwdriver.(Phillips head #1 screwdriver)

2.

Insert the attached Battery.

Do not remove the battery from its tube casing.

3.

Check the rubber packing condition and close
 the cover in the same manner as when you
 opened it.
-If dirt or scratches are present on the rubber
 packing, water resistance will be reduced. 
 Please remove the dirt or replace the rubber 
 packing if there're any cuts or scratches.
-Be sure to completely close the cover.
-Make sure not to over tighten the screws.

   

Appropriate torque: 20N/cm

30N/cm

2Kgf/cm

3Kgf/cm

〕)

Notes about Battery Installation

- After inserting the battery, it may occur that nothing appears in the display 

for about 10 seconds; this is not a malfunction.

- If a new battery has been installed and nothing appears in the display, 

please remove and re-insert the battery.

- When inserting a battery, make sure no water or foreign objects get inside the case.
- Make sure that + and – are in the correct direction.

About Lithium Batteries

-  Please store the lithium battery LS14250 in a place that is 20

 or less.

-  When using lithium batteries other than LS14250 produced by SAFT, such as 

CR2, product specifications cannot be guaranteed nor can the performance of 
some functions, such as the battery warning function.

-  Please avoid using the CR2 in the following situations: 
   Using the unit in an environment below 0

 or above 60

   Exposing the CR2 to continuous vibration such as in transportation
-  When using a CR2 lithium battery, the tube is not necessary.
-  To maintain waterproof capacity, when changing batteries also change the 

rubber packing and the drying agent (silica gel). When using a CR2 lithium 
battery, please purchase the optional Maintenance Set (TR-00P1) to replace 
the rubber packing and silica gel.

Cover

Rubber Packing

  

Attached Battery 

           (with tube)

Screws

Thermistor

 ②

Fluoropolymer Coated Electrical Wire

Specialized Connector

Appearance Diagram and Part Names

Infrared Port

           Optical 
Communication Area

Sensor 

Jack

       ALM

Warning LED

Temperature Sensor TR-5106 (Supplied with TR-52i)

 Thermo Recorder TR-51i/52i will be referred to as the 

Unit

 in this manual.

Make sure that the sensor is completely 
inserted until you hear a 

click

 sound.

The TR-52i only becomes splash proof  
when the sensor has been connected.

600

Waterproof

2.2

[unit: millimeters]

Specifications

Device Name

TR-51i

TR-52i

Measurement Item

Temperature

Temperature

Number of Channels

1 Ch (Internal Sensor Type)

1 Ch (External Sensor Type)

Measurement Range

- 40 to 80

- 60 to 155

Response Time (in 90% still air)

About 35 minutes

-

Measuring Accuracy

Avg.

±

0.5

Avg. 0.3

[-20 to 80

]

Avg. 0.5

[-40 to -20

 / 80 to 110

]

Avg. 1.0

[-60 to -40

 / 110 to 155

]

Measurement  Display Resolution

0.1

Recording Intervals

Select from 15 choices:1, 2, 5, 10, 15, 20 and 30 seconds / 1, 2, 5, 
10, 15, 20, 30 and 60 minutes 

Storage Capacity

Up to 16,000 readings

Recording Start Method

Immediate Start / Programmed Start

Recording Modes

Endless / One Time

LCD Displayed Items

Measured Temperature, Recording Status, Recording Mode Infrared 
Communication Status
Battery Life Warning, Unit of Measurement, Full (Storage Capacity 
FULL), Unconnected Sensor 
Measurement Range Exceeded, Upper/Lower Limit Exceeded

Communication Interfaces

Optical / Infrared Communication

Infrared Communication

IrPHY 1.2 low power

Communication Time

When downloading a Unit at full storage capacity:

Optical Communication : about 25 seconds (TR-50U) / about   

   150 seconds (other devices)

Infrared Communication: about 55 seconds (TR-57DCi)

Power

Lithium Battery (LS14250

Battery Life

*1

About 4 years ( 2 years if it

s been selected to 

Permit

"

 infrared 

communication)

Waterproof Capacity

Immersion proof

Splash proof

*2 

Dimensions

H62

×

W47

×

D19mm

excluding protrusions and sensor part

Weight

About 54g 

(including battery)

About 55g (including battery / excluding sensor)

Operating Environment

- 40 to 80

Data Collection Devices

Communication Ports

TR-50U2 / TR-50U

Data Collectors: TR-57DCi / 57U, RTR-57U

(*1)  Battery life depends upon the measuring environment, recording interval, and quality of the battery being 

used. When infrared communication function is enabled, battery life may be shortened if the unit is used under 
the inverter type fluorescent lighting.                            

(*2) This is the waterproof capacity of the data logger with the sensor connected.
                                                                                          *The specifications listed above are subject to change without notice.

Temperature Sensor TR-5106

Sensor Temperature Durability

- 70 to 180

Responsiveness

Thermal Time Constant: 
Approx. 30 sec. (in air) / Approx. 4 sec. (in agitated water)

Response Time (90%): 
Approx. 80 sec. (in air) / Approx. 7 sec. (in agitated water) 

TR-51i

Internal Temperature Sensor Type

Immersion proof

TR-52i

External Temperature Sensor Type

Splash proof

Package Contents

Data Logger and Common Accessories for 

TR-51i and TR-52i

Package Contents

Data Logger, Temperature Sensor TR-5106, and Common 

Accessories for TR-51i and TR-52i.

Common Accessories for TR-51i and TR-52i 

Lithium Battery LS14250 with Tube, Strap, Introductory Manual (including Warranty)

  

Introductory Manual

Thank you for purchasing this product.  

Carefully read and fully understand these instructions before 

using this unit. 

Shimadachi 817-1, Matsumoto, Nagano 390-0852 Japan

Tel:+81-263-40-0131  Fax:+81-263-40-3152

Homepage:http://www.tandd.com/   E-mail:[email protected]

Copyright  T&D Corporation. All rights reserved.  2016. 07 16504710011 (6th Edition)

This is printed using recycled paper.

 

Cautions about the Temperature Sensor

   When using the included Sensor TR-5106, please take note of the following:

Do not bend or press the last 5 cm to the tip of the sensor, as this may

damage it.

This may cause trouble or break the wire.

If the fluoropolymer-coated section of the sensor and/or the cable has a 

defect or tear, the waterproof capacity will be lost. 

Inspect it before operation.

Insert the sensor tip to about 5cm or more to obtain on accurate temperature 

measurement.

Only use the sensor within the sensor temperature durability range.

Radio, EMC and Safety Regulations

This device complies with Part 15 of the FCC Rules.
Operation is subject to following two conditions: (1) this device may not cause harmful 

interference. and (2) this device must accept any interference received, including 

interference that may cause undesired operation.

FCC Statement

This equipment has been tested and found to comply with the limits for a Class B digital 

device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide 

reasonable protection against harmful interference in a residential installation. This 

equipment generates, uses and can radiate radio frequency energy and, if not installed 

and used in accordance with the instructions, may cause harmful interference to radio 

communications. However, there is no guarantee that interference will not occur in a 

particular installation. If this equipment does cause harmful interference to radio or 

television reception, which can be determined by turning the equipment off and on, the 

user is encouraged to try to correct the interference by one or more of the following 

measures:

  -- Reorient or relocate the receiving antenna.
  -- Increase the separation between the equipment and receiver.
  -- Connect the equipment into an outlet on a circuit different from that to which  the receiver is 
connected.
  -- Consult the dealer or an experienced radio/TV technician for help.

T

o comply with the limits for the Class B digital device, pursuant to Part 15 of the FCC Rules, this device 

must be installed in computer equipment certified to comply with the Class B limits. 
All cables used to connect the computer and peripherals must be shielded and grounded. Operation 
with non-certified computers or non-shielded cables may result in interference to radio or television 
reception.

Caution

Changes or modifications not expressly approved by the party responsible for compliance 

could void the user's authority to operate the equipment. 

Thermo Recorder TR-51i / 52i Warranty

Guarantee Period

1 year from date of purchase

Date of Purchase

Customer's name

Address

Phone No.

Distributor's name

Address

Phone No.

Object of Repair

Main Unit (excluding Input Modules and any other options.)

Method of Repair

Send in for Repair

Provisions for Free Repair

1.  If the unit does not work properly despite the fact that the customer used it properly and in line 

with the manual, the unit shall be repaired free of charge through the distributor which sold the 
unit.

2.  If the customer requests free repair because of trouble within the warranty period, bring or 

send the unit along with the warranty to the distributor.  

3.  If you have moved after purchasing, or there are difficulties contacting the distributor from 

which you purchased the unit, please contact T&D directly for service. 

4.  Free repair is not available in the following cases even though it is within the warranty period:

1. Trouble or damage was caused by careless operation, natural disaster, fire, public pollution, 

or use of a power source other than specified.

2. If repair, adjustment, disassembly or modification of the unit has been carried out by a 

person other than a T&D authorized engineer.

3. Trouble or damage was caused by transportation, movement or dropping of the unit after 

purchase.

4. Failure to submit the warranty or failure to fill in all items required in the warranty.

5.  The warranty cannot be reissued. 

This warranty only promises customers free repair within the period and conditions clarified 
in this warranty. Therefore, the customer's legal rights will not be limited by this warranty. For 
further information on repair and other service questions after the termination of the warranty 
period, contact your distributor.

                                                   

Reviews: