About Menus
[Rec.Method: Instant / Average / Ave.Fine]
Select the recording method to be used. For information about the
sampling interval when recording average values, see the attached
document “Sampling Interval for Recording Average Values”.
Instant (Value): Records the instantaneous value at the set recording
interval.
Average (Value): Records the value as the average of the measure-
ments taken during the set recording interval.
Ave.(Value) Fine: This records the average value for measurements at
even shorter measurement interval.
IMPORTANT
-
When recording average values, the first recorded value is invalid because an
average cannot be calculated.
-If the Preheat function has been set to ON, then only recording of instantaneous
values can be selected.
-Recording in the “Ave.Fine” method consumes more power than other methods,
resulting in a shorter battery life.
[50-60 Hz Filter : ON / OFF ]
Switch ON/OFF the filter function to eliminate 50-60Hz noise.
*
This setting will affect both the shortest possible recording interval and battery
life.
[Rec. Interval : (user-specified time)]
Select from 23 choices: 2, 5, 10, 20, 50, 100, 200, 500 msec. / 1, 2, 5, 10,
15, 20, 30 sec. / 1, 2, 5, 10, 15, 20, 30, 60 min. (default setting is 10 min)
However, available choices will differ depending upon the combina
-
tion of other settings.
[Rec. Channel : ]
Select the channel(s) for recording. It is possible to select to record
using one channel or a combination of channels from Ch1 - Ch4. The
number of channels which can be selected will differ depending upon
the combination of other settings.
When the recording interval is less than 1 sec, the number of
channels that can be recorded simultaneously varies depending
upon the combination of settings made for 50-60Hz Filter and
Recording Interval.
Recording Intervals
2ms
5ms
10ms 20ms 50ms 100ms 200ms 500ms
1sec or
more
60min
50-60Hz
Filter
OFF
1ch
2ch
3ch
4ch
50-60Hz
Filter
ON
1ch
2ch
3ch
4ch
IMPORTANT
-When you select “Auto” as the measurement range, recording intervals of less than
1 second will not be available to be set. If you wish to set the recording interval to
less than 1 second, the measurement range must be set to other than “Auto”.
-
It is not possible to make different recording interval settings for each channel.
[CH1 Settings - CH4 Settings]
The following menus can be set for each channel.
[Meas. Range : ±300mV / ±1.5V / ±6V / ±24V / Auto*]
Select the desired measurement range.
* Auto is available only when the detailed settings is set to ON.
[Scale Conv. : ON / OFF]
Select whether or not to use scale conversion.
[Eq.]
Shows the scale conversion equation set in MCR for Windows (see
p.47). This equation is applied in the measurement displayed in
the LCD, trend graph displayed in the LCD, and recorded data graph
viewed on a PC.