T&D MCR-4V User Manual Download Page 17

30

31

About Menus

[Rec.Method: Instant / Average / Ave.Fine]

Select the recording method to be used. For information about the 

sampling interval when recording average values, see the attached 

document “Sampling Interval for Recording Average Values”.
Instant (Value): Records the instantaneous value at the set recording 

interval.
Average (Value): Records the value as the average of the measure-

ments taken during the set recording interval.
Ave.(Value) Fine: This records the average value for measurements at 

even shorter measurement interval.

IMPORTANT

 -

When recording average values, the first recorded value is invalid because an 

average cannot be calculated.

 -If the Preheat function has been set to ON, then only recording of instantaneous 

values can be selected.

 -Recording in the “Ave.Fine” method consumes more power than other methods, 

resulting in a shorter battery life.

[50-60 Hz Filter : ON / OFF ]

Switch ON/OFF the filter function to eliminate 50-60Hz noise.

This setting will affect both the shortest possible recording interval and battery 

life.

[Rec. Interval : (user-specified time)]

Select from 23 choices: 2, 5, 10, 20, 50, 100, 200, 500 msec. / 1, 2, 5, 10, 

15, 20, 30 sec. / 1, 2, 5, 10, 15, 20, 30, 60 min. (default setting is 10 min)             

However, available choices will differ depending upon the combina

-

tion of other settings.

[Rec. Channel : ]

Select the channel(s) for recording. It is possible to select to record 

using one channel or a combination of channels from Ch1 - Ch4. The 

number of channels which can be selected will differ depending upon 

the combination of other settings.

When the recording interval is less than 1 sec, the number of 

channels that can be recorded simultaneously varies depending 

upon the combination of settings made for 50-60Hz Filter and 

Recording Interval.

Recording Intervals

2ms

5ms

10ms 20ms 50ms 100ms 200ms 500ms

1sec or 

more

60min

50-60Hz 

Filter

OFF

1ch
2ch
3ch
4ch

50-60Hz 

Filter

ON

1ch
2ch
3ch
4ch

IMPORTANT

 -When you select “Auto” as the measurement range, recording intervals of less than 

1 second will not be available to be set. If you wish to set the recording interval to 

less than 1 second, the measurement range must be set to other than “Auto”.

 -

It is not possible to make different recording interval settings for each channel.

[CH1 Settings - CH4 Settings]

The following menus can be set for each channel.

[Meas. Range : ±300mV / ±1.5V / ±6V / ±24V / Auto*]

Select the desired measurement range.

* Auto is available only when the detailed settings is set to ON.

[Scale Conv. : ON / OFF]

Select whether or not to use scale conversion.

[Eq.]

Shows the scale conversion equation set in MCR for Windows (see 

p.47). This equation is applied in the measurement displayed in 

the LCD, trend graph displayed in the LCD, and recorded data graph 

viewed on a PC.

Summary of Contents for MCR-4V

Page 1: ...ons of the installation and operation of the MCR 4V Please make sure the following items are included in the package MCR 4V AA Alkaline Battery LR6 x 2 Copyr ight T D Corpo ration All rights reserv ed...

Page 2: ...manual cannot be reissued so please keep it in a safe place Please read the warranty and provisions for free repair carefully Compliance Information FCC Statement This device complies with Part 15 of...

Page 3: ...ate and or industrial use only It should not be used in situations where strict safety precautions are necessary such as with medical equipment or in systems directly or indirectly connected with huma...

Page 4: ...n Areas exposed to excessive smoke dust or dirt CAUTION Other Precautions Use the unit in the specified operating environment Do not use it for any purpose other than for which it was designed Condens...

Page 5: ...pping Recording by Group 23 About Menus Record Settings Menu 24 Data List Menu 26 Displayed Items when Recording Stopped 26 Displayed Items during Endless Recording 27 Memory Card Menu 28 Operation Se...

Page 6: ...tings Recording Settings Downloading Recorded Data Use T D Graph for Graph Display Data Analysis Printing and Saving in Text Main Features in Rec Settings Average Value Recording 50 60Hz Filter Increa...

Page 7: ...desired area of touch panel Press and hold hold down Touch the screen and hold finger in place for a few seconds Mainly used to switch ON OFF the power and key lock Control Wheel Operation rotate Pre...

Page 8: ...page Running on battery 6 Operation Mode Shows the operation mode status Touchpad operation enabled Control wheel operation enabled Touch Panel disabled 7 These indicate that there are more menu item...

Page 9: ...is not only possible to transfer data stored in the internal memory to an SD memory card but it is also possible to automatically backup data while recording in endless mode Operations related to the...

Page 10: ...o may result in loss of data being recorded When you wish to replace batteries while recording is in progress make sure to connect the USB cable to your PC for USB bus power If the MCR 4V runs out of...

Page 11: ...p 31 CH1 Settings CH4 Settings Also the scale conversion equation can be changed via MCR for Windows About the Trend Graph Display A trend graph is displayed for each channel of data 1 2 3 4 The line...

Page 12: ...y the Master but Group Recording cannot be started if the measurement interval is less than 100ms When using MCR 4TC as the Master and MCR 4V as the Slave The measurement interval of the Slave is dete...

Page 13: ...ge on the target connected Slave unit or use MCR for Windows by connecting each Slave unit to the PC with a USB cable Among the above Settings not controlled by the Master the Channel Name and Scale C...

Page 14: ...24V Auto 2 CH3 Settings Scale Conv OFF ON OFF CH4 Settings Eq y 1 0x 0 0 Unit V Warn Montr OFF OFF Lower Upper Up Low Uppr Limit 3 User Specified Voltage Value Lwr Limit 3 User Specified Voltage Value...

Page 15: ...51 Export All Data Back OK Delete Selctd Data Back No4 Jul 30 13 16 19 45 No3 Jul 29 13 15 42 51 Delete All Data Back On This Device Only On All Devices 1 1 When selecting Delete All Data on the Slav...

Page 16: ...r details about Group Recording see p 20 Detailed Sett ON OFF Switch ON OFF the detailed settings When detailed settings is switched to OFF the Rec Method 50 60Hz Filter and Rec Channel menus will not...

Page 17: ...combina tion of other settings Rec Channel Select the channel s for recording It is possible to select to record using one channel or a combination of channels from Ch1 Ch4 The number of channels whic...

Page 18: ...warning at the moment it exceeds the set upper or lower limit Preheat OFF User Specified Time Select whether or not to use the preheat function When using the preheat function make settings for the le...

Page 19: ...for the stop trigger are only detected at the points of input signal measurement at the point of recording even if the conditions are met they will not be detected Save Restore Sett By selecting Save...

Page 20: ...ee p 42 for details Delete Selctd Data While Processing is displayed on the LCD do not perform any other operations This process may take a long time to complete Select the data you wish to delete fro...

Page 21: ...n times the recording interval value The horizontal axis of the graph display shows a total of 100 dots 10 dots per box 10 boxes 100 dots There are ten data points dots in each divisional box Enlarge...

Page 22: ...r carry out a Quick Format see p 42 During the Auto Data Export process other operations cannot be performed except for switching the LCD display between Menu Measurements Trend Graph by pressing Canc...

Page 23: ...o confirm the contrast Check Memory You can check the memory usage of capacity and the number of recorded data Set Clock 1 Tapping will change the selection item of the display in the order of month d...

Page 24: ...or recover deleted or corrupted data Do not turn off the power of the MCR 4V during initialization Termination of the initialization process may cause damage or malfunction to the device Using a PC wi...

Page 25: ...en installed correctly Please see Help for Unit Recognition Failure in the Help menu and check the USB driver Making Settings via PC The following items are not included on the device menu and can onl...

Page 26: ...Windows By connecting the device to a PC with a USB communication cable it is possible to download recorded data to a PC 1Open MCR for Windows 2Connect the device to your computer 3Click Get List in t...

Page 27: ...cable during communication between a device and the PC Analyzing Recorded Data Open Recorded Data In the storage folder for the downloaded recorded data on your PC p 48 double click on the desired re...

Page 28: ...tantaneous values only and the preheat time is set to longer than the waiting time before recording start the first data reading will be invalid because preheating has not been completed 3Q Is it poss...

Page 29: ...ollowing The software may not have been installed correctly Try reinstalling MCR for Windows Check to see if the software was installed with Administrator rights If after checking and trying the above...

Page 30: ...From a slave unit Approx 4 minutes 30 seconds External Memory SD Memory Card SDHC Memory Card For Manual or Automatic Data Export Power AA Alkaline Battery LR6 x 2 AA Ni MH Battery x 2 AC Adaptor AD...

Page 31: ...A list of distributors can be found at tandd com purchasing Product Information Product information can be found at tandd com product Multichannel Voltage Recorder MCR 4V User s Manual 817 1 Shimadach...

Reviews: