Takk 5740 Operating Instructions Download Page 1

The 5740 Static Meter has been designed for the accurate analysis of industrial static electricity problems.  
It is important that the instrument is used correctly. Please read the following operating instructions and  

notes on the electric field which give useful guidelines when measuring static electricity.

Operating instructiOns

The earth lead should be used to avoid the charge in the human body affecting the readings. The ring end is 

clipped to the bottom of the meter; the crocodile clip should be connected to earth, such as a bare metal part  
of the machine.

(1)  Press blue button to zero display and take reading. Meter should be pointed at the object to be  

measured and reading taken when the top probe is 4" away. Please note that readings should  
not be taken close to electrical static eliminators or other generators of ionized air, which distorts  
the reading.

(2)  Press button again to "hold" reading. Press again to resume readings. Automatic shut-off after  

60 seconds.

(3)  The display shows the potential of the charge in kV to 1 decimal place. The maximum reading  

is 200kV. The polarity is positive unless a negative sign is shown on the display.

(4)  Please replace 9V PP3 battery when “Battery Low” indicator shows on display.

Maintenance

The charge plate in the top of the meter should be kept clean. It is important to keep this area free from  
dust and other contaminants, and must not be touched in operation, as this will affect the reading.

Apart from the battery compartment, there are no user serviceable parts inside the 5740. If there is a 
malfunction it should be returned to the manufacturer. The guarantee is void if the instrument is taken apart.

calibratiOn

The stability of the electronics in the 5740 means that recalibration is not usually needed unless the  

customer’s quality system requires it. TAKK offers a recalibration service. Alternatively an outside 

 

calibration service can be used.

The 5740 is calibrated according to:

 

EOS/ESD 3 Standard using a 150 x150mm charge plate

Or

 

BS 7506-2 using a 500mm x 500mm charge plate

Model 5740 static Meter

operating instructions