Using via Digital Data Communication
114
HX-compatible Command Details
This section explains the SET commands of HX based commands.
A
:Assigns Device Address
Function:
Assigns the device address for HX-S-G4 power supplies.
Format:
A* *: Set value within range
Range:
0~50
Assigned Address 0 is a global address. Assign addresses to all multi-connected HX-S-G4 Series power
supplies.
Once the global address is assigned, only Output ON/OFF Toggle SET command "OT" is valid. Further,
no Acknowledge Responses will be sent only while the global address is assigned
<Ex.>
A1 /* Assigns HX-S-G4 power supply of address 1 */
<Bad Example>
A1,OT1,A2,OT1 /* Error due to multiple address assignment commands */
AR
: Executes Alarm Reset
Function:
Executes alarm reset (Cancels generated Over Voltage/Over Current alarm)
Format:
AR* *: Set value
Set Value:
1: Resets alarms
Values other than 0 and 1 result in parameter errors (When 0, nothing is executed).
<Ex.>
A1,AR1
CL
:Resets Set Parameters
Function:
Resets the set parameters
Format:
CL* *: Set value
Set Value:
1: Resets to factory default settings
Values other than 0 and 1 result in parameter errors (When 0, nothing is executed).
<Ex.>
A1,CL1
Do not set redundant HX-S-G4 power supply addresses within the same system.
Further, only one address can be assigned per send. Assigning multiple addresses at once is not
possible.
Execute alarm reset only once cause has been resolved.
Caution
Caution
Summary of Contents for HX-S-G4 Series
Page 157: ...Special Loads 157 This page is blank...
Page 161: ...Maintenance 161 This page is blank...
Page 197: ...Specifications 197 External Diagrams 1 HX S 030 200G4 F I...
Page 198: ...Specifications 198 2 HX S 030 400G4 F I...
Page 199: ...Specifications 199 3 HX S 060 100G4 F I...
Page 200: ...Specifications 200 4 HX S 060 200G4 F I...
Page 201: ...Specifications 201 5 HX S 0500 12G4 F I...
Page 202: ...Specifications 202 6 HX S 0500 24G4 F I...
Page 203: ...Specifications 203 7 HX S 01000 6G4 F I...