Setting Operation during Alarm
Setting can be done about whether input shutoff is done or switching stop (forced output OFF) is done for this machine
at various types of alarm occurrence.
Setting Procedures
Press The Function “FUNC.” key.
The FUNC. lamp will light.
The Setting Item Number will display on the top Number Display
and the parameter will be displayed in the bottom Number
Display.
Select Item number 9 with the VOLTAGE dial.
Set Parameter to 0 1, or 2 with the CURRENT dial.
Parameter = 0 [Default value]:
When any alarm occurs, switching is stopped (forced output
OFF).
Parameter = 1:
At occurrence of TRIP input and serious failures (failures of
this machine), the input switch is turned off and input shutoff
is done.
Parameter = 2:
At occurrence of TRIP input, serious failures (failures of this
machine), over voltage (OVP), or over current (OCP), the
input switch is turned off and input shutoff is done.
Press the FUNC. key again to finalize settings.
Summary of Contents for HX-S-G2 Series
Page 165: ...Special Loads 165 This page is blank...
Page 169: ...Maintenance 169 This page is blank...
Page 211: ...Specifications 211 External Diagrams 1 HX S 030 200G2 F I...
Page 212: ...Specifications 212 2 HX S 030 400G2 F I...
Page 213: ...Specifications 213 3 HX S 060 100G2 F I...
Page 214: ...Specifications 214 4 HX S 060 200G2 F I...
Page 215: ...Specifications 215 5 HX S 0120 50G2 F I...
Page 216: ...Specifications 216 6 HX S 0120 100G2 F I...
Page 217: ...Specifications 217 7 HX S 0400 15G2 F I...
Page 218: ...Specifications 218 8 HX S 0400 30G2 F I...
Page 219: ...Specifications 219 9 HX S 0500 12G2 F I...
Page 220: ...Specifications 220 10 HX S 0500 24G2 F I...
Page 221: ...Specifications 221 11 HX S 01000 6G2 F I...
Page 222: ...Specifications 222 12 HX S 01000 12G2 F I...
Page 223: ...Specifications 223...
Page 239: ......