Calibration
182
Offset/Full-Scale Calibration of Output Current with External Analog Input
Output Current Control with External Voltage (Isolated External Voltage)
1. Perform load with only the shunt resistor.
Disconnect the load wires and the remote sensing
wires.
Connect the shunt to the output terminal. Connect
the digital multimeter to the terminal located on the
shunt.
Turn the POWER switch ON and OUTPUT OFF in
advance.
Note: Large current flows through
the wiring material.
Use wires which will not produce
a voltage drop below 0.8V.
Calibration Circuit
Power Supply
+
-
+
-
V
5
18
External
Control Connector
External
Voltage
0 to 10V
Insulation
Control Connector
7
6
2. Set the external voltage to 0.010V and the
constant voltage setting value to [ xV ] and set
the output to ON.
30-200
30-400
60-100
60-200
[xV]
1.00V
1.00V
1.00V
1.00V
500-12
500-24
1000-6
1000-12
[xV]
1.0V
1.0V
1V
1V
120-50
120-100
400-15
400-30
[xV]
1.0V
1.00V
1.0V
1.0V
For how to set, refer to Output Current Control with
External Voltage. (page
In the case of isolation, refer to Output Current
Control with Isolated External Voltage.
(page156)
3. Press the FUNC. key.
4. Turn the VOLTAGE dial to display the setting item
number 30 in the Number Display (voltage).
* At that time, turn off the Number Display (current).
5. Adjust so that the output current becomes [ yA ]
with the CURRENT dial.
30-200
30-400
60-100
60-200
[yA]
0.20A
0.40A
0.10A
0.20A
DMM display
value
0.050mV
0.040mV
0.025mV
0.050mV
500-12
500-24
1000-6
1000-12
[yA]
0.012A
0.024A
0.006A
0.012A
DMM display
value
0.040mV
0.040mV
0.030mV
0.040mV
120-50
120-100
400-15
400-30
[yA]
0.05A
0.10A
0.015A
0.03A
DMM display
value
0.0125mV
0.025mV
0.050mV
0.050mV
*DMM display value is the specification of the shunt resistor
The value when had you use.
6. Press the FUNC. key to settle the offset calibration
value of output current control with external
voltage (Isolated external voltage).
Summary of Contents for HX-S-01000-12G2
Page 165: ...Special Loads 165 This page is blank...
Page 169: ...Maintenance 169 This page is blank...
Page 211: ...Specifications 211 External Diagrams 1 HX S 030 200G2 F I...
Page 212: ...Specifications 212 2 HX S 030 400G2 F I...
Page 213: ...Specifications 213 3 HX S 060 100G2 F I...
Page 214: ...Specifications 214 4 HX S 060 200G2 F I...
Page 215: ...Specifications 215 5 HX S 0120 50G2 F I...
Page 216: ...Specifications 216 6 HX S 0120 100G2 F I...
Page 217: ...Specifications 217 7 HX S 0400 15G2 F I...
Page 218: ...Specifications 218 8 HX S 0400 30G2 F I...
Page 219: ...Specifications 219 9 HX S 0500 12G2 F I...
Page 220: ...Specifications 220 10 HX S 0500 24G2 F I...
Page 221: ...Specifications 221 11 HX S 01000 6G2 F I...
Page 222: ...Specifications 222 12 HX S 01000 12G2 F I...
Page 223: ...Specifications 223...
Page 239: ......