Calibration
176
Output Voltage Control with External Resistance(A)
1. Disconnect the load wires and the remote
sensing wires. Connect the digital multimeter to
the output terminal.
Turn the POWER switch ON and OUTPUT OFF in
advance.
Power Supply
+
-
V
+
-
Digital
Multimeter
Calibration Circuit
7
20
External
Control Connector
External Resistance
0
to ∞Ω
2. Set the external resistance to infinity (open) and
the constant current setting value to [ yA ] and
set the output to ON.
30-200
30-400
60-100
60-200
[yA]
1.0A
1.0A
1.0A
1.0A
500-12
500-24
1000-6
1000-12
[yA]
0.10A
0.10A
0.010A
0.10A
120-50
120-100
400-15
400-30
[yA]
0.10A
1.0A
0.10A
0.10A
On how to set, refer to Output Voltage Control (A)
with External Resistance.(page145)
3. Press the FUNC. key.
4. Turn the VOLTAGE dial to display the setting item
number 28 in the Number Display (voltage).
* At that time, turn off the Number Display (current).
5. Adjust so that the output voltage becomes "0.000V"
with the CURRENT dial.
6. Press the FUNC. key to settle the offset calibration
value of output voltage control with external
resistance (A).
Summary of Contents for HX-S-01000-12G2
Page 165: ...Special Loads 165 This page is blank...
Page 169: ...Maintenance 169 This page is blank...
Page 211: ...Specifications 211 External Diagrams 1 HX S 030 200G2 F I...
Page 212: ...Specifications 212 2 HX S 030 400G2 F I...
Page 213: ...Specifications 213 3 HX S 060 100G2 F I...
Page 214: ...Specifications 214 4 HX S 060 200G2 F I...
Page 215: ...Specifications 215 5 HX S 0120 50G2 F I...
Page 216: ...Specifications 216 6 HX S 0120 100G2 F I...
Page 217: ...Specifications 217 7 HX S 0400 15G2 F I...
Page 218: ...Specifications 218 8 HX S 0400 30G2 F I...
Page 219: ...Specifications 219 9 HX S 0500 12G2 F I...
Page 220: ...Specifications 220 10 HX S 0500 24G2 F I...
Page 221: ...Specifications 221 11 HX S 01000 6G2 F I...
Page 222: ...Specifications 222 12 HX S 01000 12G2 F I...
Page 223: ...Specifications 223...
Page 239: ......