The setup menu in your TAGARNO FHD microscope provides you
with 2 entrances to setup your microscope exactly as you like it.
 

 

Access the Setup menu by pressing the round center

 

button on the control box for approx. 1 second.

The 5 possibilities are:
1. Camera settings   

2. General settings 

3. Applications  

4. Information

5. On/off
 

 

To navigate in the set-up menu, use the multi function

 

buttons on the control box.

 

When applicable, sliding the button to the right, will activate  

 

the feature. Sliding the button left will deactivate the  

 

functionality.

Exit setup menu

You can always exit the setup menu by pressing the center button on 
the control box for 1 second or by choosing Exit in the Setup menu. 

Same procedure is valid regardless of the menu option accessed. 
 

 

1. CAMERA SETTINGS

 

1. White balance calibrate

Perform an automatic white balance adjustment in order to get
colors displayed correctly on screen. There must be a white object

present in the camera’s field of view when the white balance is being

adjusted. A white balance calibration is made each time the camera
is turned on, to adjust to the given light conditions.

NB! If the camera’s field of view does not contain a white object,

when the unit is turned on, the white balance will be incorrect and it
will be necessary to perform the white balance adjustment process
as described previously.

2. Optimal working distance

This setup works in 3 steps and needs to be carried out on a flat 
surface. Take 3 zoom levels into consideration to perform as precise 

measurements as possible. The magnification settings prior to 

performance will be restored. 

3. Closeup Lens

Indicate which lens is mounted on your TAGARNO microscope,

to get the right magnification levels displayed in the On Screen

Display when zooming in and out.

4. Start magnification

Choose the start magnification level you want your TAGARNO

microscope to automatically set when turning your system on.

5. Lock magnification

Lock your magnification level, so your microscope is always
working at your start magnification level and you are not able to

manually zoom in/out.

6. High contrast

This feature can be activated to visually enhance the edges of an 
object.Deactivated by default. 

7. Wide dynamic range

When this option is selected and auto exposure mode is active, the 
camera uses individual exposure settings for dark and light areas. This 
option can be useful to reduce glare when inspecting blank objects. 

8. Picture flip

Select if you want to display your image on screen as placed below
the camera or flipped 180 degrees horizontally.

9. IR filter

Turn Infrared filter off when using Infrared ring light. 
To use regular ring lights turn the Infrared filter on.

10. White balance calibration at LED light change

Switch on if you want an automatic white balance calibration to be 
performed when LED light is changed. 

Switch off if you do not want an automatic white balance calibration to  
be performed when LED light is changed

8.  SETUP MENU

SETUP MENU (1/5)

Summary of Contents for FHD UNO

Page 1: ...ION 7 8 SETUP MENU 10 9 SYSTEM UPDATE 15 MANUAL TAGARNO FHD UNO 10 LICENSE ACTIVATION 17 11 TECHNICAL SPECIFICATIONS 18 12 SYSTEM REQUIREMENTS 21 13 MAINTENANCE 22 14 WARRANTY 22 15 LICENSE AGREEMENT 22 16 ERGONOMIC RECOMMENDATIONS 22 17 DECLARATION OF CONFORMITY 23 18 COMPLIANCE STATEMENTS 24 Go to tagarno com productmanuals to find manuals and tutorial videos ...

Page 2: ...system before unplugging when possible WARNINGS The product is a digital magnifying system consisting of a camera unit PCBs mechanical parts and a power supply The product is intended for marketing worldwide and is designed for manual visual inspection INTENDED USE Read all safety information before you use the product Please pay attention when you see a warning label on the product This product i...

Page 3: ...3 Rod washers screws USB memory stick Manual Camera arm USB 3 0 A B cable Power supply Hexagon tool 4 YOU HAVE RECEIVED OPTIONS XPLUS FHD control box XKEY control box Base XY table HDMI HDMI cable ...

Page 4: ... video here www tagarno com installuno Assemble base and rod with washer and screw Lock the camera arm and screw the washer on top of the rod Place the camera arm on the rod The complete microscope with base ASSEMBLING BASE 1 2 5 ASSEMBLING ...

Page 5: ...Assemble XY table block and rod with three screws Lock the camera arm and screw the washers on top of the rod Place the camera arm on the rod The complete microscope with XY table ASSEMBLING XY TABLE 2 2 ...

Page 6: ...ol box D SUB 9 Pin Female Control box Output 3 3V 6A Max Use only control boxes supplied by TAGARNO 4 Output HDMI Standard Type A Monitor HDMI Out 1080p60 5 Ethernet RJ 45 Ethernet wall socket Ethernet LAN 100BASE TX 1000BASE T 6 USB 3 0 output USB 3 0 Type B Socket Computer USB 3 0 Device Self powered 7 ESD Snap Socket 10mm ESD Grounding point ESD Grounding point Only black models 8 Power supply ...

Page 7: ...ens you are using Adjust the height of the camera arm according to the mounted lens The three coloured rings on the rod mark the correct focus height for a 3 4 or 5 lens respectively Blue Red Black 7 OPERATION Lens Ring colour Lens to object distance 3 Blue 333 mm 13 1 4 Red 250 mm 9 8 5 Black 200 mm 7 9 Lens to object distance with a 10 lens is 100 mm 3 9 ...

Page 8: ... reduce the iris instead of the exposure time to improve the depth of view Manual focus Turn autofocus off by pushing both zoom buttons plus and minus on the control box at the same time until you hear a beep You can now adjust focus manually with the help of the zoom buttons Push this button on the control box to adjust focus if you wish to focus on something very close Push this button on the co...

Page 9: ... short pressing the button Factory default directs to the Ruler application with both horizontal og vertical ruler activated Multi Function Buttons When the Setup menu is deactivated you are able to use the Multi Function Buttons to control the iris gain and exposure time Press the center button to switch between auto and manual exposure mode 491190 Folie til XPLUS FHD 13 06 2017 Zoom Push this bu...

Page 10: ...up works in 3 steps and needs to be carried out on a flat surface Take 3 zoom levels into consideration to perform as precise measurements as possible The magnification settings prior to performance will be restored 3 Closeup Lens Indicate which lens is mounted on your TAGARNO microscope to get the right magnification levels displayed in the On Screen Display when zooming in and out 4 Start magnif...

Page 11: ...ew the image at the center of the display for 5 seconds Image preview can now be disabled which will display a more discrete text box for 3 seconds 2 GENERAL SETTINGS 1 Ruler color Set the ruler color to your prefered color choice to separate it from your displayed object Choose between 6 different colors YELLOW BLUE RED BLACK WHITE AND GREEN 2 Lock presets When this feature is enabled users are p...

Page 12: ... while Monitor auto detect is enabled will disable the Monitor auto detect function 8 System units Choose to display metrics or imperial units when using the microscope 9 Numeric format Here you can select the numeric format based on your preference Choose between 1 234 56 1 234 56 or 1 234 56 10 System language The following languages are available as system language Chinese Danish English French...

Page 13: ...s you to perform precise measurements such as line radius or angle measurements directly from the microscope 3 Ruler Turn the Ruler application on horizontally vertically or both directions and get a ruler displayed on top of your live image on screen 4 Watermark Create an info field on captured images with an image prefix time stamp horizontal field of view microscope model plus serial number and...

Page 14: ...the system to the factory default settings all your personal settings will be erased including calibrations and images 5 Format USB storage This TAGARNO FHD device is designed to only support USB storages with a FAT32 formatted file system If your USB storage does not currently have this file system use this function to format the USB storage A second pop up will ask you to confirm your selection ...

Page 15: ... stick FAT32 formatted 3 If the file s is in a ZIP folder unzip the files and make sure that only the correct installation file s or previous system update files appear on the memory stick Note File names always have the following format tagarno_fhd_system_vx_xx image tagarno_fhd_system_vx_xx system tagarno_fhd_system_vx_xx update x_xx denoting the system version The installation file s is identic...

Page 16: ...scribed above entering system update STEP 5 VERIFY INSTALLATION 1 Verify the system update by checking that the Application version displayed in the Setup menu Info System information matches the version of the system update file STEP 6 ACCESS THE APPLICATION If a new application is available a new menu point will appear in the Setup menu under pt 3 Applications showing the new application Note To...

Page 17: ... TAGARNO application you might need to activate a unique license key file to be able to access and use the application To activate a license key on your TAGARNO microscope follow the steps below Note The license can only be paired with the microscope which has the serial number stated in your license The license can not be removed and transfered to another microscope STEP 1 DOWNLOAD LICENSE KEY FI...

Page 18: ... A Socket USB 3 0 output Type B Socket Control box D SUB 9 Pin Female ESD Snap Socket 10mm Monitor output HDMI Standard Type A Power supply DC Jack Ethernet RJ 45 Power requirements Line voltage 100 240V 10 Line frequency 50 60Hz AC current consumption typ 150mA 100V 70mA 240V DC current consumption typ 1 3A 12V Environmental conditions Temperature Storage 5 to 60o C 23 to 140o F Operation 5 to 40...

Page 19: ... 3 0 output Type B Socket Control box D SUB 9 Pin Female ESD Snap Socket 10mm Monitor output HDMI Standard Type A Power supply DC Jack Ethernet RJ 45 Power requirements Line voltage 100 240V 10 Line frequency 50 60Hz AC current consumption typ 150mA 100V 70mA 240V DC current consumption typ 1 3A 12V Environmental conditions Temperature Storage 5 to 60o C 23 to 140o F Operation 5 to 40o C 41 to 104...

Page 20: ...e calibration Choose image format Choose keyboard layout Type image prefix Select foot switch configuration Activate Ruler Set optimal working distance Choose ruler color Set time Watermark Optional Choose a close up lens Activate file share mode Set date Verification lines Optional Turn wide dynamic range on off Turn high contrast on off Image preview Image comparison Optional Set start magnifica...

Page 21: ... RAM nVidia GeForce GT 740M with 2GB RAM Monitor resolution 1920x1080 FULL HD 1080P 25 30HZ OS Desktop Laptop Windows 7 8 or 10 with DirectX 11 Memory 4GB CPU Intel Core Dual 3 0GHz or i3 2 4GHz 3rd generation named 3xxx Intel Core i3 3 2GHz 3rd generation named 3xxx Hard Disc Space 1GB required SSD type recommended Connections USB 3 0 xHCI host controller Intel chipset recommended Integrated Grap...

Page 22: ...elivery and will be free from defects in material and workmanship for a period of 24 months 2 years from date of invoice from Supplier This warranty covers a The replacement of defective parts b All labour costs to exchange defective parts in product c Complete functions test of product before return to end user s site d Return transport costs of product from TAGARNO to end user s site The warrant...

Page 23: ...es that the product listed above consisting of a camera unit and a 12V power supply is in compliance with the following European directives 2006 25 EU 2014 30 EU 2014 35 EU By conforming to the following harmonized standards IEC 62471 2008 EN 61326 1 2013 IEC 61010 1 2010 The declaration is issued by MANUFACTURER TAGARNO A S Risk Group Exempt Class A Industrial Electromagnetic Environment IECEE CB...

Page 24: ...d can radiate radio frequency energy and if not installed and used in accordance with the instruction manual may cause harmful interference to radio communications Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense INDUSTRY CANADA COMPLIANCE STATEMENT CAN ICES 3 A NMB 3 A T...

Reviews: