Lucid Series RF Signal Generator
Benchtop User Manual
Rev. 1.4
22
•
Frequency
–
Set the pulse frequency in Hz.
•
Source
:
⧫
Internal
–
Use the screen modulation parameters.
⧫
External
–
Use a pulse source connected to the generator’s MODULATION IN connector
located on the rear panel. The Generator will accept modulating signals between ±1 V (2 V p-
p) amplitude.
•
Push the
Mod On/Off
button on the device front panel to start the modulation and then
RF On/Off
button to output the signal.
3.2.5
Pattern Sequence
Select on the device display the Modulation tab, and then click the PATTERN button to show the
pattern sequence parameters. You can set a sequence of pulses according to the list of pulses where
each step in the list defines a pulse Time On and Time Off time and the number of Loops.
Note
•
The
PATTERN
button is only available if the device has the PAT option installed.
Figure 3.6 Pattern Sequence
Define a patter according to the steps below.
1.
NEW
–
Click the button to create a new Pattern.
2.
LOAD
–
Click the button to select a file to load.
3.
SAVE
–
Click the button to save the Pattern in a JSON (
JavaScript Object Notation)
format. You
can save to a new name (valid name is only numbers) or overwrite an old file. You can select to
store the data on the SD card.
4.
DEL
–
Click the button to select a file to delete.
5.
+STEP
–
Click the button to add a new step.