DHR/AR Series Environmental Test Chamber Getting Started Guide
Page 16
Installing the ETC Assembly on the AR Series
Follow the instructions below to install the ETC on an AR-G2 or AR 2000ex.
Mounting the Test Chamber
1
Turn on the rheometer and move the rheometer head up to the maximum height. Use the
Head UP
button, located on the instrument control panel.
2
Fit the air-bearing clamp to the rheometer.
3
Turn off the power to the rheometer control box.
4
Ensure that the two top screws (
A
and
B
in the figure below) are fitted with washers and are located in
place—but make sure that they are almost totally unscrewed (two turns in).
Figure 9
Mounting screws.
5
Open the ETC oven and then use the handles on the oven doors to lift it onto the two top screws. Lightly