RSA-G2 Getting Started Guide
Page 48
Leveling the RSA-G2
The RSA-G2 is leveled using the feet, or levelers (shown below), located on each corner. Note that the
right-rear leveler is fixed in position;
do not attempt to adjust this leveler
.
Clockwise
rotation of the leveler extends it, raising the instrument at its location.
Counterclockwise
rotation of the leveler retracts it, lowering the instrument at its location.
Figure 30
RSA-G2 leveler.
Follow the instructions below to perform this procedure.
1
Make sure that the air is on and the motor is unlocked.
2
Place the supplied bull’s eye level on top of the motor mount.
Figure 31
Bull’s eye level placed on lower geometry.
3
Retract all three (3) levelers by turning them counterclockwise until the instrument is lowered as far as
possible.
Clockwise
Counterclockwise