RSA-G2 Getting Started Guide
Page 30
Instrument Specifications
The table found below contains the technical specifications for the RSA-G2.
Table 4: RSA-G2 Technical Specifications
Item/Area
Specifications
Test station
Depth 56 cm (22 in.)
Width 46 cm (18 in.)
Height 104 cm (39.37 in.)
Power supply enclosure
Depth 48 cm (19 in.)
Width 32 cm (12.5 in.)
Height 34 cm (13.5 in.)
Oven supply enclosure
Depth 48 cm (19 in.)
Width 32 cm (12.5 in.)
Height 11 cm (4.5 in.)
Weight
:
Test station
Power supply enclosure
Oven supply enclosure
104.5 kg (< 230 lbs)
12.7 kg (28 lbs)
6.8 kg (15 lbs)
Power
:
Test station
Power supply enclosure
Oven supply enclosure
From power supply:
90 – 264 VAC, 16 amps max
180 – 264 VAC, 47 – 63 Hz, 13 amp max
Operating environmental conditions
Temperature: 15–30°C
Relative humidity: 5–80% (non-condensing)
Installation Category II
Pollution degree 2
Maximum altitude: 2000 m (6560 ft)
Minimum clearances of six inches from top and all sides,
and eight inches from temperature control options.
The degree of protection for this instrument according to
EN 60529 is IP20.
TEMPERATURE SYSTEMS
Forced Convection Oven (FCO)
–150 to 500°C (lower limit dependent on cooling device)
FCO Camera Viewer
Optional
Precision
+0.05°C
Stability
+0.1°C
Summary of Contents for RSA-G2
Page 1: ......
Page 9: ...Forced Convection Oven with tension clamp Technology 8...
Page 13: ...IMMERSION TESTING I CLAMPING SYSTEMS...
Page 15: ...DETA I DIELECTRIC THERMAL ANALYSIS ACCESSORY DETA...
Page 29: ...tainstruments com 2019 TA Instruments All rights reserved L50027 001...
Page 30: ...Verso Filler Page z z...
Page 31: ...n getting Started guide Revision D Issued July 2022...
Page 99: ...Issued November 2013 Supplement to Getting Started Guide TA Instruments Manual Supplement...
Page 100: ...Page 2 TA Instruments Manual Supplement This page is intentionally blank...
Page 113: ...Page 1 Page 1 RSA G2 Solids Analyzer Revision G Issued May 2022 Site Preparation Guide...