Page 2: ...Press ON ZERO key 0 2 C A L 5 Press TARE key C A L 0 0 6 Press TARE key to make the SECOND digit flash C A L 0 0 7 Press ON ZERO key to change the flashing digit to 1 C A L 0 1 8 Press TARE key Display will show full scale capacity 3 0 0 0 9903II Calibration Instructions P a g e1o f2 A Division of T E Bouton Company Inc ...
Page 3: ... is stable You will hear 3 beeps Full capacity weight value will flash in display window 3 0 0 0 12 Remove the weight from the platform and press any key C A L 0 1 13 Press UNITS key C A L 0 0 14 Press TARE key 0 2 C A L 15 Turn scale OFF 16 Re adjust silver switch located on lower portion of the main board to the original position 17 Turn the scale OFF and close it up 9903II Calibration Instructi...