SZBJ BM920 Series Operating Instruction Download Page 1

 

 

BM920 

Operating Instruction 

 

I    Overview

   

The  complete  machine  circuit  design  of  BM920  series  takes  large-scale 

integrated  circuit  and  A/D  converter  as  the  core,  provided  with  global  function 
overloading protection to measure direct current voltage and current, A.C. voltage 
and  current,  resistor  and  current,  resistor,  capacitor,  frequency  temperature, 
diode  forward  voltage  drop,  transistor  hFE  parameters  and  circuit  on  and  off  , 
Infrared remote control signal detection

 

and

 

Identification of live wire    etc. 

 

II Safety rules and attentions

 

The  instrument  of  model  BM920  is  in  line  with  IEC1010-1  CAT

  1000V 

standard. Before use, please read the instruction manual carefully.   

It  is  not  allowed  to  use  with  the  rear  cover  not  ready;  otherwise,  it  may  cause 

danger.   

Before  use,  check  insulation  layer  of  the  pen-shape  meter  shall  be  perfect  and 

free of damage and broke wires.   

The range switch shall be set at the correct range position. 

The  input  signal  can  not  be  beyond  the  specified  limits  to  avoid  electric  shock 

and damaging the instrument.   

Shift of the range switch can not be changed during measuring. 

Measure  potential  difference  of  the  public  terminal  COM  and  the  ground  “ 

 

 

“ can not exceed 600V to avoid electric shock.   

For  testing  voltage  higher  than  DC50V  and  AC36Vrms,  be  careful  to  avoid 

electric shock.   

When  the  LCD  displays  “

”,  electric  quantity  of  the  battery  is  not  sufficient, 

and the battery shall be replaced to ensure normal operation of the instrument. 

Replacement of fuse in the instrument shall employ the same specifications.   

The instrument shall be calibrated once a year at least.   

III Performance 

1. 

Insufficient electric quantity indication: “

”.   

2. 

Maximum display: 1999

3 1/2 bit

.   

3. 

Automatic  switching  off:  automatically  cut  off  power  supply  after 
switching on the machine about 15 minutes. 

4. 

10A jack: without fuse.   

5. 

mA jack fuse: 0.2A/250V 

Self-recovery fuse. 

6. 

Battery in the unit: 9V      NEDA or 6F22 or equivalence.   

7. 

Environmental conditions: 

1

  Working temperature: 0

40

 

  RH<75%   

2

  Storage temperature: -10

50

 

RH<85 % 

IV Technical indexes

 

Precision ± (a % r figures

  Warranty period: 1 year 

Guaranteed temperature precision: 23

±5

 RH<70 % 

1. Direct current voltage 

Range 

Resolution 

Measurement error 

200mV 

0.1mV 

2V 

1mV 

20V 

10mV 

200V 

100mV 

±

0.5%+5d

 

1000V 

1V 

±

0.8%+5d

 

Input impedance: 200mV and 2V shift is 1MΩ, and others are 10MΩ.   
Overloading protection: 200mV and 2V range is 250V, and others are 
  direct or A.C. PEAK 1000V.   

2. A.C. voltage 

Measurement error 

Range 

Resolution 

BM9205 

BM9208 

200mV 

0.1mV 

±

1.2%+5d

 

2V 

1mV 

20V 

10mV 

200V 

100mV 

 

±

0.8%+5d

 

 

±

0.8%+5d

 

700V 

1V 

±

1.2%+7d

 

Input impedance: 200mV and 2V range is 1MΩ, and others are 10MΩ.   
Frequency scope: 40Hz

400Hz (200V and 700V ranges are 40Hz

100Hz).   

Overloading  protection:  200mV  and  2V  ranges  are  250V,  and  others  are  direct 
current or alternative current peak value1000V.   
Display: average value (sine wave valid value calibration). 

3. Direct current 

Range 

Resolution 

Measurement error 

20uA 

0.01μA 

±

1.2%+5d

 

20mA 

10μA 

±

0.8%+5d

 

200mA 

100μA 

±

1.2%+5d

 

10A 

10mA 

±

2%+5d

 

Overloading protection: 0.2A/250V fuse and 10A range without fuse.   

Maximum input current: 10A (input time no more than 10 seconds). 

Measuring voltage drop: the full range is DC200mV.   

4. Alternative current 

Measurement error 

Range 

Resolution 

BM9205 

BM9208 

20mA 

10μA 

±

1.2%+5d

 

200mA 

100μA 

±

1.8%+5d

 

10A 

10mA 

±

3%+7d

 

Overloading protection: 0.2A/250V with fuse, and 10A range without fuse 
Maximum input current: 10A (input time no more than 10 seconds) 
Measuring voltage drop: the full range is 200mV.   
Frequency scope: 40Hz

400Hz.   

Display: average value (sine wave valid value calibration)   

5. Resistance 

Range 

Resolution 

Measurement error 

200Ω 

0.1Ω 

2KΩ 

1Ω 

20KΩ 

10Ω 

200KΩ 

100Ω 

2MΩ 

1KΩ 

 
 

±

0.8%+5d

 

20MΩ 

10KΩ 

±

1%+5d

 

200MΩ 

100KΩ 

±

5%+10d

-10d 

Overloading protection: 220V valid value 
Open circuit voltage:

1V (200MΩ range is 2.8V) 

About 10 displayed in case of short circuit for 200MΩ range: actual measurement 
value = display value -10 
For example: the measured standard resistance is 100MΩ, the instrument display 
is 101.0 and the real value is 101.0 - 10 = 100.0 

6. Capacitance Cx 

Range 

Resolution 

Measurement error 

20nF 

10pF 

200nF 

100pF 

2μF 

1nF 

 

±

2.5%+20d

 

 

200μF 

100nF 

±

5%+10d 

Overloading protection: 36V    DC or peak value

 

7. Frequency F 

Measurement error 

Range 

Resolution 

BM9208 

2KHz 

1Hz 

20KHz 

10Hz 

±

1.5%+5d

 

Notes: valid value of the tested frequency signal: 300mV~50V 
            Overloading protection: 220V valid value 

8. Temperature

 

Measurement error 

Range 

Resolution 

BM9208 

-20

400

 

1

 

±

0.75%+4d

 

401

1000

 

1

 

±

1.5%+15d

 

Notes: model K thermal coupler is used (temperature probe).   

9. Crystal triode hFE testing 

Range 

Notes 

Testing conditions 

hFE 

Displayed scope 

0

1000β 

Ibo≈10μA, 

Vce≈2.8V 

10. Diode and buzzing continuous conductance testing

 

Range 

Notes 

Testing conditions 

 
 

 

Display  forward  conductance 
round value of diode 

Forward direct current about 

1mA 

Backward direct current voltage 

about 2.8V 

 

 

When 

the 

conductance 

resistance  is

  about  50Ω, 

buzzer  in  the  machine  works, 
and 

round  value  of  the 

resistance is displayed. 

Open circuit voltage about 2.8V 

Overloading protection: 220V valid value 

11. Infrared remote control signal detection

 (

)    and 

Identification of live wire     

(▲)    See instructions 

V Operation instructions

 

      Before use, pay attention to the symbol beside the jacket of the testing pen 

”, which reminds you the tested voltage or current not beyond this value. In 

addition, before use, the range switch shall be set at the switch which you want 
to set. 

1. D.C. voltage measurement

 