CHAPTER 10
Troubleshooting
10-14
Sysmex XE-5000 Instructions for Use
R
e
vi
se
d A
p
ri
l 20
07
Error messages:
FCM TD Therm Sens ERR
Status:
Not Ready
Category:
Analysis Error
Probable Cause
1. One of the thermal sensors for optical detector block has
broken.
Corrective Action
1. Turn the power OFF, then ON again. If the error persists,
contact your Sysmex technical representative.
Error messages:
Env Therm Sens ERR
Status:
Not Ready
Category:
Analysis Error
Probable Cause
1. One of the thermal sensors for detecting outside air
temperature has broken.
Corrective Action
1. Turn the power OFF, then ON again. If the error persists,
contact your Sysmex technical representative.
Error messages:
IMI Reac Temp High HPC
IMI Reac Temp Low HPC
Status:
Not Ready
Category:
Analysis Error
Probable Cause
1. IMI reaction temperature is outside the regulation range for
HPC analysis mode.
Corrective Action
1. IMI reaction temperature stabilization:
Press the HELP key on the Main Unit panel keypad. Select
“OK” on the function menu to execute IMI temperature
stabilization sequence. In this case, it is possible to perform
analysis in normal mode even if the error persists.
Error messages:
Env Temp High HPC
Env Temp Low HPC
Status:
Not Ready
Category:
Analysis Error
Probable Cause
1. The ambient temperature in which the instrument is used is
outside the the regulation range for HPC analysis mode.
Corrective Action
1. Check the ambient temperature:
Keep surrounding temperature within the range 20-30°C, in
which HPC analysis mode is usable. In this case, it is
possible to perform analysis in normal mode even if the
error persists.
Error messages:
IMI Temp Stab. ERR
Status:
Not Ready
Probable Cause
1. Before the IMI reaction temperature stabilized, the transition
to HPC mode was processed, and the HPC sleep
restoration process was completed.
Corrective Action
1. IMI reaction temperature stabilization:
Press the HELP key on the Main Unit panel keypad. Select
“OK” on the function menu to execute IMI temperature
stabilization sequence. In this case, it is possible to perform
analysis in normal mode even if the error persists.
Summary of Contents for XE-5000
Page 2: ......
Page 54: ...CHAPTER 5 Before Using 5 4 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Page 96: ...CHAPTER 6 Operation 6 42 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Page 128: ...CHAPTER 8 Calibration 8 10 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Page 258: ...CHAPTER 12 Warranty 12 2 Sysmex XE 5000 Instructions for Use Revised April 2007 Blank page ...
Page 262: ...CHAPTER 13 Index Revised April 2007 13 4 Sysmex XE 5000 Instructions for Use Blank page ...