Universal Battery Charger II Instructions for Use DePuy Synthes 19
The device must be unplugged before it is cleaned. To
clean the charger, wipe it off with a clean, soft and lint-
free cloth dampened with deionized water and dry prior
to reprocessing.
Then wipe the charger with a new, clean, soft and lint-
freeclothdampenedwithaminimumof70% alcohol-
based disinfectant for thirty (30) seconds. A dis in fectant
that is VAH (Verbund für Angewandte Hygiene) listed,
EPA (Environmental Protection Agency) registered or
locally recognized is recommended. This step has to be
repeated two (2) additional times using a new, clean,
soft and lint-free cloth dampened with a minimum 70%
alco hol-based disinfectant each time. Please follow the
safety instructions given by the manual of the disinfec-
tant manufacturer which is selected.
Whenever it is cleaned, the device should be checked to
ensure it is working properly and is not damaged.
Maintenance of the device is not required.
If there are any faults, please send the device to a
DePuy Synthes representative (see next section).
Precautions:
•
Danger of electric shock! Unplug before cleaning.
•
Do not sterilize, wash, rinse, drop or apply force to
the UBC II. This will destroy it with possible
secondary damage.
•
If necessary, clean the contacts in the charger bays
using utmost care.
•
Do not to spray the contacts or touch both contacts
at the same time with the damp cloth due to dan-
ger of short circuiting. Avoid contact of the device
with fluids.
Warning:
Make sure that no solution enters the
device.
Care and Maintenance
Cleaning
Summary of Contents for Universal Battery Charger II
Page 1: ...Universal Battery Charger II Instructions for Use...
Page 2: ......