Universal Battery Charger II Instructions for Use DePuy Synthes 17
Immediately recharge batteries and the power module
after each use.
Any Colibri/SBD batteries (532.003, 532.033) or Battery
Power Line battery (530.620) that are not used should
always be stored in the activated charger (maintenance
charge). This guarantees that the batteries are always
fully charged and ready to use.
It is not necessary to store Trauma Recon System power
module (05.001.202), Battery Power Line II battery
(530.630) and Colibri II/SBD II battery (532.103) in the
charger. Once charged they can be stored outside the
charger without any noticeable charge difference due to a
very low self-discharging rate of Lithium-based batteries.
Warning:
Unusable or faulty batteries/power mod-
ules must not be reused and should be disposed of
in an environmentally friendly manner and in
accordance with national regulations. For further
information, refer to the section ‘Disposal’ in this
IFU.
Operating the Device
Storing Batteries