14 DePuy Synthes Instructions for Use Universal Battery Charger II
2. Battery Power Line II and Colibri II / SBD II
batteries
The charger enables Battery Power Line II and Colibri II /
SBD II batteries (530.630, 532.103) to be checked. It will
beindicatedifthebatteryperformanceissufficientorif
the battery needs to be replaced.
Due to the very low self-discharging rate of Lithium-
based batteries a refresh function is not required.
When required, the check function is started manually,
as outlined below.
After the battery is inserted, the symbol
lights up
yellow. To check the battery, press the button with the
exclamation mark foratleast2seconds(fig.5)until
the symbol lightsupyellow(fig.6).Thedevicethen
carries out the process. The symbol is yellow through-
out this time.
Precautions:
•
To recharge batteries normally, do not press the
exclamation mark button .
•
Do not remove the battery from the charger bay as
long as the symbol is yellow (fig. 6). Wait until
the process has ended and the symbols
or
light up. Only then is the battery status clearly
assessed.
Fig. 5
Fig. 6
Operating the Device
Checking and Refreshing Batteries