Universal Battery Charger II Instructions for Use DePuy Synthes 13
Completion of the process is indicated as follows:
•
Symbol
isgreen(fig.3):Batteryhasbeensuccess-
fully refreshed, checked and charged.
•
Symbol isred(fig.4):Eitherthebatteryisfaultyor
performanceisinsufficient.Thebatterymustbedis-
posed of.
The entire process (refreshing and checking battery
status) takes around 10 hours and should only
be carried out if there is enough time to do so.
A battery can be charged, checked or refreshed inde-
pendently in each charger bay.
Precautions:
•
Checking the battery status and refreshing the
battery have an impact on the battery. If this is
carried out frequently the lifespan of the battery
can be affected.
•
The process is interrupted if there is a power cut
or a switch to the emergency power supply and
will then have to be restarted.