Load washing machine basket

Place articles in the specially designed trays for machine
washing supplied by Synthes. Ensure that all cannulations
(handpiece and attachments), if applicable, are positioned
vertically, i.e., in an upright position as shown.

Note:

Never immerse the handpiece, lid or attachments in

an ultrasonic bath or any other liquid as this could decrease
the service life of the system. 

Mechanical cleaning cycle parameters

Minimum

Cycle

time (minutes)

Temperature/water

Type of detergent

Rinse

2

Cold tap water

N/A

Prewash

1

Warm tap water (40ºC)

Neutral pH enzymatic cleaner

Cleaning

2

Warm tap water (45ºC)

Neutral pH enzymatic cleaner

Rinse

5

DI (Deionized) water

N/A

Thermal disinfection

5

Hot DI water (93ºC)

N/A

Dry

40

90ºC

N/A

9

Inspect devices

Remove all devices from washing basket. Inspect the 
cannulations, coupling sleeves, etc. for visible soil. If 
necessary, repeat the manual preclean/mechanical cleaning
cycle. Please refer to Synthes Battery Powered Equipment 
Lubrication Chart after cleaning and prior to sterilization.                             