Technical Specifications
6
Synthes
Anspach Electric Systems
Instructions for Use
Table 1: Emission
The Synthes Anspach High Speed Electric System is composed of Handpiece models eMax2, eMax2 Plus, E12, and E-Saggital;
Electric Consoles models SC2000, SC2000U, SC2100, SC2101, and SC2102; Foot Control models E-FP, E-FP-DIR, E-FP-DIR/IRR,
EPLUS-FP, EPLUS-FP-NS, EMAX2-FP, EMAX2-FP-NS; Hand Controls EMAX2-HC and E12-HC.
Guidance and manufacturer’s declaration – electromagnetic emissions
The Synthes Anspach High Speed Electric System is intended for use in the electromagnetic environment specified below. The
customer or the user of the Synthes Anspach High Speed Electric System should assure that it is used in such an environment.
Emission test
Compliance
Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1
The Synthes Anspach High Speed Electric System uses RF energy only for its
internal function. Therefore, its RF emissions are very low and are not likely
to cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class A
The Synthes Anspach High Speed Electric System is suitable for use in all
establish ments, other than domestic establishments and those directly
connected to the public low-voltage power supply network that supplies
buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
Complies
0x6.001.507_AD 13.06.14 11:11 Seite 6