Minimal Access Attachment Bearing Sleeves
Follow the Attachment cleaning protocol for the bearing
sleeves. The Minimal Access Attachment Bearing Sleeves are
non-repairable. Reuse until bearing wear is detected and
then replace. No bearing sleeve service or repair will be pro-
vided.
Dissection Tools
Warning:
Disposable. Single use only.
Electric Console and Foot Control
Caution:
DO NOT IMMERSE OR STERILIZE! DO NOT allow
liquid to enter console or foot control.
1. Disconnect all power from console.
2. Clean console, foot control, and irrigation pump by
wiping with non-abrasive cloth and disinfectant or mild
detergent and water after each case.
3. Dry thoroughly with non-abrasive cloth.
Electric High Power Sagittal Saw
Caution:
DO NOT IMMERSE! DO NOT allow liquid to enter
Electric High Power Sagittal Saw.
1. Disconnect from console.
2. Wipe with non-abrasive cloth and disinfectant or mild
deter gent and deionized, distilled, or purified water.
3. Dry thoroughly with non-abrasive cloth.
Handpieces
Warning:
Wear eye protection.
Caution:
DO NOT IMMERSE! Do not insert brush into hand-
piece housing. Do not use saline solution for cleaning. Do
not use cleaners containing chlorinated phenols of any con-
centration. Using cleaners/disinfecting agents containing
chlorinated phenols will result in premature hose failure.
Prior to sterilization, wipe with a clean, water-dampened
cloth and mild detergent (neutral pH) until no evidence of
soil is found.
Silicone Spray or Instrument Milk may be added to lubricate
the handpiece knurled knob.
Silicone spray may be used on the hose to prolong its flexi -
bility and to facilitate foreign particle removal.
The handpiece should be tested by running it with an attach-
ment and dissection tool prior to every use for one minute to
ensure proper operation. If the distal tip and main body of
the handpiece are uncomfortably hot to the touch, return
the instrument to Synthes Anspach for servicing.
Caution:
Handpiece must not be exposed to ingress of wa-
ter or to severe physical trauma; degradation of unit function
and/or performance may occur.
Sterilization Trays
Sterilization trays can be cleaned with deionized, distilled, or
purified water and a mild detergent.
Mechanical Cleaning
For mechanical cleaning instructions, see information sup-
plied with appropriate cleaning and sterilization basket for
instruments being cleaned.
Warning:
Transmissible Spongiform Encephalopathies (TSE).
Synthes Anspach will not authorize or accept the return of
products that directly contact patients or are contaminated
with a patient's body fluids who is suspected or confirmed
with a Transmissible Spongiform Encephalopathies/
Creutzfeldt-Jakob Disease (TSE/CJD) diagnosis. Synthes
Anspach recommends that all Anspach products used on a
patient confirmed with a TSE/CJD diagnosis be incinerated.
Anspach dissecting tools used on a patient suspected of
TSE/CJD diagnosis must be incinerated.
Contact your Sales Representative for replacement of prod-
uct incinerated under this policy or for temporary equipment
while original equipment is quarantined. Contact the Synthes
Anspach Customer Service Department regarding TSE/CJD
contamination for additional information.