The microSaws were designed for small bone dissection.
The microSaw Driver provides direct connection to all
Anspach electric handpiece systems. The Driver also enables
rapid exchange of saw head styles through a unique quick-
disconnect system.
Keyless Driver Installation
DRIVER Keyless Driver
Note:
Instructions are written such that distal end of hand-
piece and attachment are pointed away from user.
Keyless Driver Installation
1. Ensure knurled knob on handpiece is in “RUN” (
) posi-
tion.
2. Insert drive shaft of keyless driver into handpiece and ro-
tate clockwise until locked. Gently twist back and forth to
ensure it is locked into position.
eMax Keyless Driver Removal (DRIVER)
1. Move handpiece knurled knob to “SAFE” position.
2. Rotate DRIVER counterclockwise and remove.
DRIVER Duty Cycle
30 Sec ON, 30 Sec OFF for 8 cycles.
The recommendations for times of use for the DRIVER at-
tachments have been determined under average load and
worst case ambient air temperature of 85 °F (29 °C).
DRIVER
microSaws and Small Attachments
Assembly
30
Synthes
Anspach Electric Systems
Instructions for Use
0x6.001.507_AD 13.06.14 11:12 Seite 30