Bearing Sleeve and Dissection Tool Assembly
(20° Angle Driver or Straight Driver)
MA-10S Bearing Sleeve, 10 cm Straight
MA-10C Bearing Sleeve, 10 cm Curved
MA-15S Bearing Sleeve, 15 cm Straight
MA-15C Bearing Sleeve, 15 cm Curved
MA-15ST Bearing Sleeve, 15 cm Straight Taper
MA-19ST Bearing Sleeve, 19 cm Straight Taper
1. Rotate knurled knob on driver to line up arrow with
“TUBE”.
2. Fully insert desired bearing sleeve into distal end of driver
by lining up black line on bearing sleeve with black line
on driver.
3. Rotate knurled knob to line up arrow with “CUTTER”.
Check security of bearing sleeve by pulling distally.
4. Insert dissection tool into distal tip of bearing sleeve.
Rotate dissection tool slowly until it is fully seated. Rotate
knurled knob on driver to line arrow with “RUN”.
Note:
Additional force is required to seat dissection tool into
curved bearing sleeve.
5. Check security of dissection tool by pulling distally on
dissection tool head.
6. Bearing sleeve and dissection tool are secure.
Dissection Tool Disassembly
1. Rotate knurled knob on driver to line up arrow with
“CUTTER”. Remove dissection tool from distal end of
attach ment.
Bearing Sleeve Adjustment
1. Rotate knurled knob on driver to line up arrow with
“ADJUST”.
2. Pull bearing sleeve either 3 mm distally or proximally to
desired exposure.
3. Rotate knurled knob on driver to line up arrow with “RUN”.
Bearing Sleeve Disassembly
1. Rotate knurled knob on driver to line up arrow with
“TUBE”. Remove bearing sleeve distal end of driver.