Note:
Instructions are the same for straight or angled drivers
and for straight or curved bearing sleeves in both 10 cm and
15 cm lengths. The four words labeled on the drivers mean
the following:
CUTTER
– Insert or remove dissection tool.
RUN
– Attachment and dissection tool are now functional.
ADJUST
– Bearing Sleeve can be moved 3 mm forward
or back.
TUBE
– Insert or remove bearing sleeve.
Warning:
The Minimal Access Attachments are to be used
only for bone dissection applications. Excessive force or side
loads during use can cause rapid temperature increase at the
distal end of the attachment. Always use copious irrigation
with this attachment.
Driver Installation
MA-D20 20° Angle Driver
MA-DRIVER Minimal Access Straight Driver
Note:
Instructions are written such that distal end of hand-
piece and attachment are pointed away from user.
eMax Driver Installation
1. Ensure handpiece knurled knob is in “RUN” (
) position.
2. Insert drive shaft of driver into distal end of handpiece,
push and rotate to right until it stops. Listen for a click as
handpiece knurled knob moves up to seat next to driver.
3. Gently twist back and forth to ensure driver is locked into
position.
MA-D20
MA-DRIVER
Minimal Access Attachment and
Dissection Tool Assembly
Anspach Electric Systems
Instructions for Use
Synthes
27
0x6.001.507_AD 13.06.14 11:12 Seite 27