Standard Attachments and Dissection Tool Assembly
Craniotome and Specialty Attachments
CRANI-P 6.5 cm Pediatric Craniotome
CRANI-A 6.5 cm Adult Craniotome
CRANI-L 7.5 cm Large Craniotome
CRANI-A-R 6.5 cm Rotating Adult Craniotome
ADG 7.2 cm Adjustable Drill Guide
CDA 6.8 cm Controlled Depth Attachment
Warning:
It is recommended that a craniotome not be used
in conjuction with Hand Control. Excessive rotational force
on hand control may move it resulting in handpiece shutting
down.
Note:
Instructions are written such that distal end of hand-
piece and attachment are pointed away from user.
1. Pull handpiece knurled knob towards “LOAD” (
) posi-
tion and insert dissection tool into distal end of hand-
piece. Rotate dissection tool slowly until dissection tool is
fully seated.
2. Release handpiece knurled knob into “RUN” (
) position.
Note:
Handpiece is now fully functional.
3. Pull dissection tool to ensure proper engagement.
4. Slide attachment over distal end of handpiece. Pull attach-
ment towards handpiece, and rotate attachment
to right approximately one-quarter turn until fully seated.
Handpiece knurled knob moves up to seat next to attachment.
5. Attachment and dissection tool are secure.
Disassembly
1. Move handpiece knurled knob to “SAFE” position.
2. Rotate attachment to left approximately one-quarter turn,
and remove from distal end of handpiece.
3. Remove dissection tool from distal end of attachment.
CRANI-P
CRANI-A
CRANI-L
CRANI-A-R
ADG
CDA
24
Synthes
Anspach Electric Systems
Instructions for Use
0x6.001.507_AD 13.06.14 11:12 Seite 24