Standard Attachments and Dissection Tool Assembly
Handpiece
Knurled Knob
Handpiece
Distal End
Arrows and
Symbols
SHORT
MEDIUM
LONG
MIA16
SHORT-HD
MEDIUM-HD
LONG-HD
XL-HD
eMax Straight Attachments
Straight Attachments
SHORT 5 cm Short Attachment
MEDIUM 8 cm Medium Attachment
LONG 11 cm Long Attachment
MIA16 16 cm Minimally Invasive Attachment
SHORT-HD 5 cm Heavy Duty Short Attachment
MEDIUM-HD 8 cm Heavy Duty Medium Attachment
LONG-HD 12.4 cm Heavy Duty Long Attachment
XL-HD 20.2 cm Heavy Duty Extra Long Attachment
Note:
Instructions are written such that distal end of hand-
piece and attachment are pointed away from user.
The arrows, symbols, and printing on the distal end of the
handpiece housing (“RUN” (
), “LOAD” (
) and “SAFE”)
refer to locking and unlocking the attachment and dissection
tool onto the handpiece.
1. Move handpiece knurled knob to “RUN” (
) position.
2. Slide attachment over distal end of handpiece. Pull
attachment towards handpiece, and rotate attachment
to right approximately one-quarter turn until fully seated.
Handpiece knurled knob moves up to seat next to attach-
ment.
3. Pull handpiece knurled knob towards “LOAD” (
) posi-
tion and insert dissection tool into distal end of attach-
ment. Rotate dissection tool slowly until fully seated.
4. Release handpiece knurled knob into “RUN” (
) position.
Note:
Handpiece is now fully functional.
5. Pull dissection tool to ensure proper engagement.
6. Attachment and dissection tool are secure.
Disassembly
1. Move handpiece knurled knob to “SAFE” position.
2. Remove dissection tool from distal end of attachment.
3. Rotate attachment to left approximately one-quarter turn,
and remove from distal end of handpiece.
22
Synthes
Anspach Electric Systems
Instructions for Use
0x6.001.507_AD 13.06.14 11:12 Seite 22