System Assembly
1. Plug the hospital grade AC power cord into the AC power
inlet on the back of the console. Plug the opposite end of
the power cord into a standard, Hospital Grade grounded
wall outlet. Connect to supply mains with protective earth
only.
2. Insert foot control cord connector into Foot Control 1 con-
nector port. (A second foot control cord may be con-
nected into Foot Control 2 connector port on applicable
consoles.) Foot control cord connector is keyed. Align foot
control cord connector with port on front of console. Do
not push foot control cord connector into console connec-
tor port when out of alignment.
3. Plug handpiece hose connector into Handpiece 1 connec-
tor port located on front of console. (A second handpiece
could be connected into Handpiece 2 connector port on
applicable consoles.) Handpiece hose connector is keyed.
Align handpiece hose connector with connector port on
front of console. Do not push handpiece hose connector
into console connector port when out of alignment.
4. Activate console by depressing power switch located on
the back of the console to the “I” position. Appropriate
Face Panel Light Emitting Diodes (LED) illuminate, and a
beep sounds.
5. For operating instructions see Console Face Panel Opera-
tion and Foot Control Operation below.
Note:
Handpiece is fully functional at this time.
Anspach Electric Systems
Instructions for Use
Synthes
13
0x6.001.507_AD 13.06.14 11:12 Seite 13