columns is the Dew Point temperature in °F (or °C). The chiller’s temperature setpoint must be set above
the dew point temperature.
For example, if the air temperature is 85 °F (29 °C) and the relative humidity is 60%, then the dew point
temperature is 70 °F (21 °C). Adjust the chiller’s temperature setpoint to 72 °F (22 °C) to prevent
condensation from forming inside the laser.
Cooling guidelines
Refer to the following dew point temperature table for a range of air temperature and relative humidity
values. Remember that the laser’s coolant temperature must be set above the dew point temperatures
shown in the chart. They should not exceed the environmental specifications located in the General
Specifications within this chapter of this manual.
Air-condition the room or the enclosure containing the laser.
Install a dehumidifier to reduce the humidity of the enclosure containing the laser.
Table 4-3 Dew point temperatures.
Summary of Contents for p250
Page 1: ...ENGINEERED BY SYNRAD p250 Laser User Manual...
Page 15: ...NOMENCLATURE CONTINUED PAGE 15 Nomenclature Continued Figure 1 2 Anatomy of a model number...
Page 27: ...EUROPEAN UNION EU REQUIREMENTS ROHS COMPLIANCE PAGE 27 Figure 2 3 p250 Declaration Document...
Page 40: ...RF POWER SUPPLY PAGE 40 Table 4 3 Dew point temperatures Continued...
Page 63: ...INPUT OUTPUT SIGNALS PAGE 63 Table 4 4 User I O pin descriptions Continued...
Page 84: ...P250 GENERAL SPECIFICATIONS PAGE 84 Table 4 5 p250 general specifications Continued...
Page 87: ...P250 OUTLINE MOUNTING DRAWINGS PAGE 87 Figure 4 47 p250 packaging instructions...
Page 93: ...OPERATIONAL FLOWCHART PAGE 93 Operational flowchart Figure 5 1 p250 operational flowchart...
Page 94: ...STATUS LED S PAGE 94 Status LED s Table 5 2 p250 Input Output LED Status Signals...
Page 105: ...INDEX PAGE 1 This page is intentionally left blank...