Editing plate measurements
To change the Reject Small Particles filter setting for an individual plate result:
1. Select the Edit Result tab for the plate result – see
2. Drag the
slider; click the arrow buttons at the ends of the slider to adjust the slider by a single
step; or press anywhere on the slider bar to set the slider to that position.
The plate will be remeasured with the new setting.
3. Check that the real colonies are still being counted while the small particles you want
to reject are being excluded, and repeat Step 2 if the result is still not satisfactory.
Note
You may find it helpful to zoom the image – see
4. Save the change to the result (see
Changing the Sensitivity setting for a result
When you create a batch you can adjust the detection Sensitivity – see
page 30. After you have accepted the batch, you can also change the Sensitivity setting in
Measurement mode
before
taking a measurement (see
Test measurement and adjusting
This section shows you how to change the Sensitivity setting for an individual plate
after
it has
been measured.
To change the detection Sensitivity setting for an individual plate result:
1. Select the Edit Result tab for the plate result – see
2. Drag the
slider; click the arrow buttons at the ends of the slider to adjust the slider by a single
step; or press anywhere on the slider bar to set the slider to that position.
The plate will be remeasured with the new setting.
3. Check that the real colonies are being counted and the background and any debris are
being excluded, and repeat Step 2 if the result is still not satisfactory.
Note
You may find it helpful to zoom the image – see
4. Save the change to the result (see
Splitting colonies in a result
When you create a batch, you can choose whether to use the colony splitter to split
overlapping colonies so that they are counted separately – see
, page 30. You
can also change the colony splitter setting in Measurement Mode
before
taking a
measurement (see
Test measurement and adjusting settings
aCOLyte3
User Manual
85
Summary of Contents for SYNBIOSIS aCOLyte3
Page 2: ......
Page 6: ......
Page 12: ...Getting started 6 aCOLyte3 User Manual...
Page 22: ...Capturing images 16 aCOLyte3 User Manual...
Page 66: ...Creating a new batch 60 aCOLyte3 User Manual...
Page 96: ...Working with results 90 aCOLyte3 User Manual...
Page 108: ...Configuring aCOLyte3 102 aCOLyte3 User Manual...