Chapter 14. Alarms and Notifications
14.1. Alarm Messages and Responses
14.1.1. There are five different alarm levels: Very High, High, Medium,
Low and Very Low. If an alarm occurs while the Driver is
supporting the patient, refer to
Table 14-1
for alarms and their
corrective actions. For alarms observed while the Driver is not
supporting a patient, contact SynCardia for additional information.
If more than one alarm occurs, the highest priority alarm message
is shown first in the Alarm Bar.
Very High, High, Medium and Low Alarms are accompanied with
an audible notice between 70-80 decibels. They vary in
occurrence and frequency as shown in
Table 14-1.
Very low
alarms do not have an audible notice.
The graphical user interface software requires two minutes to fully
load after the power is turned on with the key switch. No Driver
alarms are active during this time.
Table 14-1 – Alarm Messages and Responses
Alarm
Message
Visual Indicators
Audible Signal
Corrective Action
Dual
Compressor
Malfunction
Alarm Bar: Flashing Red
Alarm Indicator LED:
Flashing Red
10 beeps every 2½
seconds
70-80 dBA
Replace Driver with a SynCardia
approved backup Driver.
Notify SynCardia at
1-866-771-9437
Very Low
Left Cardiac
Output
Alarm Bar: Flashing Red
Alarm Indicator LED:
Flashing Red
10 beeps every 2½
seconds
70-80 dBA
Make sure the Driveline is connected
and has no kinks.
If alarm persists, replace Driver with a
SynCardia approved backup Driver.
Notify SynCardia at
1-866-771-9437
Very Low
Right Cardiac
Output
Alarm Bar: Flashing Red
Alarm Indicator LED:
Flashing Red
10 beeps every 2½
seconds
70-80 dBA
Make sure the Driveline is connected
and has no kinks.
If alarm persists, replace Driver with a
SynCardia approved backup Driver.
Notify SynCardia at
1-866-771-9437
System
Malfunction
Alarm Bar: Flashing Red
Alarm Indicator LED:
Flashing Red
10 beeps every 2½
seconds
70-80 dBA
Replace Driver with a SynCardia
approved backup Driver.
Notify SynCardia at
1-866-771-9437
Summary of Contents for Companion 2 Driver System
Page 2: ...SynCardia Systems LLC Companion 2 Driver System Operator Manual Page 2 of 132...
Page 4: ...SynCardia Systems LLC Companion 2 Driver System Operator Manual Page 4 of 132...
Page 8: ...SynCardia Systems LLC Companion 2 Driver System Operator Manual Page 8 of 132...
Page 14: ...SynCardia Systems LLC Companion 2 Driver System Operator Manual Page 14 of 132...
Page 16: ...SynCardia Systems LLC Companion 2 Driver System Operator Manual Page 16 of 132...
Page 18: ...SynCardia Systems LLC Companion 2 Driver System Operator Manual Page 18 of 132...
Page 32: ...SynCardia Systems LLC Companion 2 Driver System Operator Manual Page 32 of 132...
Page 56: ...SynCardia Systems LLC Companion 2 Driver System Operator Manual Page 56 of 132...
Page 66: ...SynCardia Systems LLC Companion 2 Driver System Operator Manual Page 66 of 132...
Page 96: ...SynCardia Systems LLC Companion 2 Driver System Operator Manual Page 96 of 132...
Page 118: ...SynCardia Systems LLC Companion 2 Driver System Operator Manual Page 118 of 132...
Page 130: ...SynCardia Systems LLC Companion 2 Driver System Operator Manual Page 130 of 132...
Page 132: ...SynCardia Systems LLC Companion 2 Driver System Operator Manual Page 132 of 132...