TimeAnalyzer 7x00 User’s Guide
098-00208-000 Revision A – February, 2010
Chapter 4 Operating
E1 Measurements
Figure 4-7.
Live TimeWatch TDEV Data
E1 Measurements with External Client
Setup
In addition to the connections shown in
, on page 91 (except
connection from TP500 module to the network switch), make the following
connections to the TimeWatch module ports:
REF IN Port
:
Connect the REF IN port to Port 1 on the I/O card of
the TP5000 probe module using a BNC-to-mini BNC
cable
MEASURE IN Port
Connect the MEASURE IN port to the E1 output of the
external client using the appropriate cable.
LAN
Connect the LAN port to the internal Ethernet switch
using Ethernet cable
Procedure
1. Connect E1 or 2M signal from TP5000 module to Reference port of TimeWatch
module. The default output from Port 1 of the TP5000 probe module is 2048 kHz.
2. Connect E1 output from the external client to Measurement port of the
TimeWatch probe module.