Chapter 2 Using the 5071A
Monitoring Status
Operating and Programming Manual
21
Monitoring Status
Use the following procedures to check the 5071A‘s operational status and selected
internal parameters.
To Select Parameters
Press
Shift
, then
7
(INFO): display shows instrument status.
Select the reading you want to check then use the cursor keys to examine the
instrument’s status.
To Interpret Displayed Results
The front-panel LCD display shows the current status (or value) of the selected
parameter. Refer to Table 2-2 below for typical internal-parameter operating ranges
and definitions.
Table 2-2. Internal Operating Parameters
Parameter
Description
Typical Range
Pump
Ion pump current
0 - 40 µA
Ovens
Cesium beam tube and VCXO Oven
monitor
0 to 10V, and
-5 to -10V
Ctrl
VCXO control voltage
-95 to 95%
C-Field
C-field control current
10.0 to 14.05 mA
Ampl
Microwave signal strength for CBT
(% of maximum)
0 to 100%, and
0 to 100%
E_Mult
CBT electron multiplier voltage
1000 to 2553V
Gain
Frequency control loop gain
(% of maximum)
14.4 to 58%
Power
Voltage source
AC, DC, or Battery
To Print a Status Report
Use the following procedure to print a status report when a printer is connected to the
rear-panel RS-232C serial port (see page 27).
1
Select the INFO menu.
2
Select the PRINT option and press
Enter
.
Summary of Contents for 5071A
Page 2: ......
Page 28: ...Preface xxvi Operating and Programming Manual ...
Page 29: ...____________________________________________ Chapter 1 Getting Started Getting Started ...
Page 32: ...Chapter 1 Getting Started The 5071A at a Glance 4 Operating and Programming Manual ...
Page 33: ...Chapter 1 Getting Started The 5071A at a Glance Operating and Programming Manual 5 ...
Page 40: ...Chapter 1 Getting Started Performing Basic Tasks 12 Operating and Programming Manual ...
Page 41: ...____________________________________________ Chapter 2 Using the 5071A Using the 5071A ...
Page 97: ...____________________________________________ Chapter 5 Command Reference Command Reference ...
Page 119: ...Chapter 5 Command Reference Description of SCPI Commands Operating and Programming Manual 91 ...
Page 136: ...Chapter 5 Command Reference Error Messages Codes 108 Operating and Programming Manual ...
Page 137: ...____________________________________________ Chapter 6 Specifications Specifications ...
Page 145: ...____________________________________________ Chapter 7 Installation Installation ...
Page 162: ...Chapter 7 Installation Cesium Beam Tube Replacement 134 Operating and Programming Manual ...
Page 186: ...Appendix A Performance Tests 5071A PERFORMANCE TEST 158 Operating and Programming Manual ...
Page 187: ...____________________________________________ Appendix B Glossary Glossary ...
Page 190: ...Appendix B Glossary 5071A Glossary 162 Operating and Programming Manual ...