![Symmetricom 5071A Operating And Programming Manual Download Page 168](http://html.mh-extra.com/html/symmetricom/5071a/5071a_operating-and-programming-manual_1419614168.webp)
Appendix A Performance Tests
5071A Operational Verification
140 Operating and Programming Manual
5071A Operational Verification
Power-On Self-Tests and Servo Lock
1
Inspect the 5071A for any damage (see the section titled Initial Inspection in
Chapter 7 for detailed inspection information).
2
Ensure that the ac-power setting, ac-line fuse (rear panel), and supplied power
cord are appropriate for the ac-power source. (Refer to Chapter 7, ―Installation,‖
for detailed information.)
3
Now, apply power to the instrument by connecting ac-power source to the
instrument via the ac power cord. The amber Attention light is lit indicating
normal power-up sequence and the following messages are displayed on the LCD
in sequence:
The message ―Waiting for Stabilization‖ may appear. If so, it could stay on for
up to 20 minutes before the following messages appear.
a.
Warming up
(this indicates self-test passed successfully)
b.
Lowering E_mult voltage
c.
Setting Osc. Control
d.
Setting RF amplitude
e.
Setting E_mult voltage
f.
Logging signal levels
g.
Setting C-field
h.
Locking servo loops
i.
Operating normally
4
After about 15 minutes (typically), the Attention (amber) light goes out and the
Continuous Operation (green) light flashes. This indicates that all servo loops
have locked and the instrument is operating normally.
5
Press
Shift
, then
5
(Utilities).
LCD display shows RESET.
6
Press
Enter
.
This resets the continuous operation circuit, causing the light to be on
steadily. Any subsequent fault will cause the Continuous Operation
light to go out or flash.
Summary of Contents for 5071A
Page 2: ......
Page 28: ...Preface xxvi Operating and Programming Manual ...
Page 29: ...____________________________________________ Chapter 1 Getting Started Getting Started ...
Page 32: ...Chapter 1 Getting Started The 5071A at a Glance 4 Operating and Programming Manual ...
Page 33: ...Chapter 1 Getting Started The 5071A at a Glance Operating and Programming Manual 5 ...
Page 40: ...Chapter 1 Getting Started Performing Basic Tasks 12 Operating and Programming Manual ...
Page 41: ...____________________________________________ Chapter 2 Using the 5071A Using the 5071A ...
Page 97: ...____________________________________________ Chapter 5 Command Reference Command Reference ...
Page 119: ...Chapter 5 Command Reference Description of SCPI Commands Operating and Programming Manual 91 ...
Page 136: ...Chapter 5 Command Reference Error Messages Codes 108 Operating and Programming Manual ...
Page 137: ...____________________________________________ Chapter 6 Specifications Specifications ...
Page 145: ...____________________________________________ Chapter 7 Installation Installation ...
Page 162: ...Chapter 7 Installation Cesium Beam Tube Replacement 134 Operating and Programming Manual ...
Page 186: ...Appendix A Performance Tests 5071A PERFORMANCE TEST 158 Operating and Programming Manual ...
Page 187: ...____________________________________________ Appendix B Glossary Glossary ...
Page 190: ...Appendix B Glossary 5071A Glossary 162 Operating and Programming Manual ...