OPERATING INSTRUCTIONS
Controls and Indicators . . . . . . . . . . . . . . . .
Rhythms for Defibrillator Training . . . . . . . . . .
ST Segment and T Wave Abnormali es . . . . . .
Defibrillator Exercises . . . . . . . . . . . . . . . .
External Pacing Exercises . . . . . . . . . . . . . .
Ba ery Saver Feature . . . . . . . . . . . . . . . .
IBP Waveform Op on . . . . . . . . . . . . . . . .
Service . . . . . . . . . . . . . . . . . . . . . . . . .
1
2
2
2
3
3
3
3
convert
When enabled, a pre-selected rhythm becomes the running rhythm,
when a defib delivers 50J or more.
1. Press
convert
key. The indicator of the running rhythm
pulses brighter.
2. Press key of the waveform to run a er the defib discharge.
3. Deliver discharge from defib set to 50J or more.
The pre-selected rhythm is now the running rhythm.
To cancel a convert opera on before its completed, press either the
convert
key or the key of the running rhythm. If a convert opera on is
started but not completed in two minutes, it will cancel automa cally.
capture
When the pacer’s current and rate are properly adjusted, paced beats
are displayed and the
PACER PULSE
indicator blinks off with each
pacer pulse detected.
Press
capture
key to select one of five pacing capture levels: 70, 80,
90 and 100 mA, or “ignore”. Brightness of
PACER PULSE
indicates
the level selected. At 70 mA, the brightness is lowest. At 100 mA,
the brightness is highest. When the
PACER PULSE
indicator is off,
“ignore” is selected and the simulator won’t respond to pacer pulses.
At power-on, the default pacing capture level is 70 mA.
ON OFF
Press to power-on and to power-off.
LOW BATT
Red indicator illuminates when ba ery
should be replaced.
DEFIB DSCHG
Green indicator illuminates for two
seconds when discharge is delivered
from defib set to 50J or more.
convert
When enabled, a pre-selected rhythm
becomes the running rhythm when a
defib delivers 50J or more.
PACER PULSE
Brightness indicates pacing capture
level selected.
capture
Selects capture current levels:
70, 80, 90 and 100 mA and “ignore”.
VF
VT-HI
VT-LO
AFIB
AFLTR
PSVT
NSR
ASYS
SINUS
SINUS
3rd
2nd I
SINUS
S BRDY
S BRDY
2nd II
1st
2nd II
ON OFF
LOW
BATT
DEFIB
DSCHG
T de P
S TACH
ANT
ST
convert
AMI
RBBB
PVC
LBBB
PACER
PULSE
capture
ANT S
ST
LAT
ST
INF
ST
IMI
AMI
PVC
LBBB
Controls and Indicators
1