5. Electrical connection
Fig. 12: Electrical connection
Transmitter
Terminal no.
Connection
Connection of the power supply
L / +24 V
Input power supply 230 V / 50 Hz, 110 V / 60 Hz (optional 24 V DC)
N / 0 V
Input power supply 230 V / 50 Hz, 110 V / 60 Hz (optional 24 V DC)
PE
Protected earth
Sensor connection
I-out1
+
Analogue
-
Analogue output -
Na
Not available
Na
Not available
Na
Not available
Na
Not available
Min. /
Max.-
relay
NO
Potential-free relay NO (Close)
C
Potential-free relay C (Common conductor)
NC
Potential-free relay NC (Open)
D-out
+
Digital
-
Digital output -
RS 485
A
RS 485 intersection data A
B
RS 485 intersection data B
GND
RS 485 intersection ground
D-in1
+
Digital intersection 1 (+)
-
Digital intersection 1 (-)
D-in2
+
Digital intersection 2 (+)
-
Digital intersection 2 (-)
Sensor
+
Power 24 V
Cable no. 1
GND
Power supply 0 V
Cable no. 2
A
RS 485 data A
Cable no. 3
B
RS 485 data B
Cable no. 4
Shield
Shield
L N PE
I-in 1 I-out 1
+ -
I-out 2 I-out 3
Alarm Relay
NO C NC
D-out
RS 485
A B
GND
D-in 1 D-in 2
Sensor
+ - A B
Shield
+ - + -
+ - + -
+ -
+ -