28
A-96.250.441 / 060619
AMU pH-Redox
Installation
To install a sensor with a shaft length of 120 mm proceed as fol-
lows:
1
Unscrew and remove the blind plug [F] from the flow cell block.
2
Carefully remove the protective cap [J] from the sensor tip. Turn
it clockwise only.
3
Rinse the sensor tip with clean water.
4
Slide the distance sleeve [C] over the sensor shaft and slightly
tighten the fixing screw.
5
Slide the union screw [D], the washer [E] and the O-ring [G]
over the sensor shaft [H].
6
Insert the sensor through the flow cell block [I] into the calibra-
tion vessel [K].
7
Tighten the union screw [D] hand-tight.
8
Remove the connector cap [A].
9
Screw the connector [B] onto the sensor.
10
Keep the protective caps in a secure place for later use.
3.5.
Flow Cell M-Flow
3.5.1
Install Swansensor pH/Redox Standard or AY
CAUTION
Fragile parts
The pH and the ORP sensors are fragile.
Handle with care.
Do not spill KCl when removing the protective cap.
Sensors
This instruction applies for both, the pH and the ORP sensor.