10
A-96.250.131 / 050220
AMI Turbitrace
Product Description
2.1. Instrument Specification
CAUTION
The measuring chamber is under the pressure of the
sample line
If the inlet pressure is too high, moisture can enter the housing
of the measuring electronics and lead to incorrect measure
-
ments and failure of the instrument.
The maximum inlet pressure of 10 bar must never be
exceeded, not even for short time periods. Take appropriate
measures to protect the instrument from pressure surges in
the sample line.
Power Supply
Voltage:
Power consumption:
100–240 VAC (±10%)
50/60 Hz (±5%)
or 24 VDC (±10%)
max. 30 VA
Electronics
Aluminium with a protection degree of IP 66 / NEMA 4X
housing
Ambient temperature:
Limit range of operation:
Storage and transport:
Humidity:
Display:
-10 to +50 °C
-25 to +65 °C
-30 to +85 °C
10–90% rel., non condensing
backlit LCD, 75x45 mm
Sample
requirements
Flow rate:
Temperature:
Inlet pressure
Outlet pressure:
5–20 l/h
1–40 °C
1–10 bar
pressure free
On-site
The analyzer site must permit connections to:
requirements
Sample inlet:
Sample outlet:
Max. Altitude:
Serto 4x6 mm
Hose nozzle, 15x20 mm
2000 m above sea level