AMI Solicon4
Program List and Explanations
A-96.250.321 / 191015
69
The relay will be switched via the Profibus input. No further param-
eters are needed.
5.3.4
Input:
The functions of the relays and signal outputs can be de-
fined depending on the position of the input contact, i.e. no function,
closed or open.
5.3.4.1
Active:
Define when the input should be active:
5.3.4.2
Signal Outputs:
Select the operation mode of the signal outputs
when the relay is active:
5.3.4.3
Output/Control:
(relay or signal output):
5.3.4.4
Fault:
5.3.4.5
Delay:
Time which the instrument waits, after the input is deactivat-
ed, before returning to normal operation.
Range: 0–6‘000 Sec
5.3.2.1
Function = Fieldbus:
No:
Input is never active.
When closed
Input is active if the input relay is closed
When open:
Input is active if the input relay is open
Cont.:
Signal outputs continue to issue the measured
value.
Hold:
Signal outputs issue the last valid measured value.
Measurement is interrupted. Errors, except fatal
errors, are not issued.
Off:
Set to 0 or 4 mA respectively. Errors, except fatal
errors, are not issued.
Cont.:
Controller continues normally.
Hold:
Controller continues on the last valid value.
Off:
Controller is switched off.
No:
No message is issued in pending error list and the
alarm relay does not close when input is active.
Yes:
Message E024 is issued and stored in the mes-
sage list. The Alarm relay closes when input is
active.