A-96.250.211 / 171121
AMI Sodium P
Maintenance
64
6.4. Maintenance of Reference Electrode
Remove the
reference
electrode
1
Unscrew and remove the electrode plug [A].
Prevent the connector from getting wet.
2
Put on the connector cap [B] onto the electrode.
3
Completely unscrew the union screw [C] from the threaded hole.
4
Remove the KCl bottle from its holder.
Remember that the bottle was punctured - do not spill KCl.
5
Remove the reference electrode from the flow cell.
Cleaning
1
Remove any iron deposits with a soft paper tissue.
2
Slip the ring-shaped sleeve [F] upwards with a turning and push
-
ing movement and allow a little electrolyte to flow out.
3
Fix the ring-shaped sleeve finger tight with a gentle turning and
pulling movement.
4
Replace or refill KCl reservoir. Use only original SWAN KCl.
Install
See
Installing the Reference Electrode, p. 27
A
B
C
D
E
F
Sensor plug
Connector cap
Union screw
Washer
O-ring
Ring-shaped sleeve
A
B
D
C
E
F