AMI Powercon
Maintenance
A-96.250.341 / 061015
53
Limits and intervals for the AMI Powercon
Procedure
The standard workflow contains following procedures:
1
Activate SWAN Quality assurance procedure
2
Pre-test
3
Connect instruments
4
Carry out comparison measurement
5
Completion of the measurement
Note:
The procedure should only be carried out through
qualified personnel.
6.8.1
Activate SWAN Quality assurance procedure
Enable quality assurance procedure at each instrument by select-
ing the quality level in menu 5.1.4.1
The corresponding submenus are then activated.
Note:
The activation is necessary the first time only.
Quality Level
max. deviation
temperature [°C]
a)
max. deviation
result [%]
min. inspection
interval
0: Off
Off
Off
Off
1: Trend
0.5 °C
10 %
annual
2: Standard
0.4 °C
5 %
quarterly
3: Crucial
0.3 °C
5 %
monthly
4: User
0 - 2°C
0 - 20%
annual, quarterly,
monthly
a) sample temperature must have 25°C +/- 5°C.