Carry out comparison measurement
The comparison measurement is menu driven. Start by selecting
Quality Assurance in menu 3.5 of the monitor AMI Oxytrace.
1
Navigate to menu Maintenance /Quality Assurance.
2
Press [Enter].
3
Follow the dialog on the Display.
4
Carry out pre test preparations
Connect instruments.
Regulate sample flow to 10 l/h us-
ing the appropriate valve.
5
Wait 10 minutes whilst measure-
ment is running.
Press [Enter] to continue.
6
Read the ppb value of the refer-
ence instrument and enter under
“Inspector” by using the [
] or
[
] keys.
7
Press [Enter] to confirm.
8
Read temperature value of the ref-
erence instrument and enter un-
der “Inspector Temp.” by using the
[
] or [
] keys.
9
Press [Enter] to confirm.
10
Press [Enter] to continue.
The results are saved in QA-
History regardless if successful
or not
3.5.5
Quality Assurance
- Carry out preparations
- Install Inspector
- Sample flow to 10 l/h
<Enter> to continue
3.5.5
Quality Assurance
Value O2
0.05 ppb
Value Temp.
25.00
C
Wait 10 Minutes
<Enter> to continue
3.5.3
Quality Assurance
Value O2
0.05 ppb
Value Temp.
25.00
C
<Enter> to continue
Inspector O2
0.06 ppb
Inspector Temp.
25.0
C
3.5.4
Quality Assurance
Value H2
0.05 ppb
Value Temp.
25.00
C
<Enter> to continue
Inspector
0.06 ppb
Inspector Temp.
25.0
C
3.5.5
Quality Assurance
Max. Dev. O2
0.5 %
Max. Dev. Temp.
0.4 °C
QA-Check succesful
Dev. O2
0.1 %
Dev. Temp.
0.4 °C
Summary of Contents for AMI Oxytrace
Page 1: ...AMI Oxytrace AMI Oxytrace QED Version 6 20 and higher A 96 250 531 300617 Operator s Manual...
Page 5: ...AMI Oxytrace QED A 96 250 531 300617 3 11 Default Values 90 12 Index 93 13 Notes 94...
Page 96: ...94 A 96 250 531 300617 AMI Oxytrace QED Notes 13 Notes...
Page 97: ...AMI Oxytrace QED Notes A 96 250 531 300617 95...