AMI Oxytrace QED
Maintenance
A-96.250.531 / 300617
49
6.6.
Zero-Verification
Swansensor Oxytrace G for the measurement of low oxygen con-
tent (< 1 ppb).
1
Calibrate the sensor according to chapter
2
Prepare a 5%-sodium sulfite solution with demineralized water.
3
Put the electrode into the sodium sulfite solution afterwards. As-
sure that there are no air bubbles in front of the sensor.
4
The measured value should now be < 1 ppb.
NOTICE:
Depending on the state of the electrode this process
can take several hours. In case of an electrode-refill, this can
take days until the measured value is lower than 1 ppb.
Summary of Contents for AMI Oxytrace
Page 1: ...AMI Oxytrace AMI Oxytrace QED Version 6 20 and higher A 96 250 531 300617 Operator s Manual...
Page 5: ...AMI Oxytrace QED A 96 250 531 300617 3 11 Default Values 90 12 Index 93 13 Notes 94...
Page 96: ...94 A 96 250 531 300617 AMI Oxytrace QED Notes 13 Notes...
Page 97: ...AMI Oxytrace QED Notes A 96 250 531 300617 95...