AMI Oxytrace QED
Maintenance
A-96.250.531 / 300617
45
6.4.
Maintenance of the Faraday electrode
1
Switch off the instrument and close the flow regulating valve.
2
Open the faraday control unit.
3
Disconnect and remove the cable from the faraday control unit.
4
Unscrew and remove the fixing sleeve (A).
5
Remove the faraday electrode from the flow cell, do not pull on
the cable.
6
Remove the washer (C) and the o-ring [D] from the electrode
body (B).
7
Unscrew the electrode tip containing the hollow electrode (G).
A
B
C
D
E
F
G
Fixing sleeve
Electrode body
Washer
O-ring
Inner electrode
O-ring
Hollow electrode
A
C
D
E
F
G
B