A-96.250.451 / 190121
AMI Deltacon Power
Installation
20
3.4. Installation of Cation Exchanger
Cation ex
-
changer bottle
The bottle containing the cation exchanger is delivered separately.
For transport, an empty bottle is installed into the bottle holder.
Install cation
exchanger
bottle
Install the cation exchanger bottle as follows:
1
Unscrew and remove the empty bottle [E] from the bottle holder
[B].
2
Fill high purity water into the cation exchanger bottle [C], until the
water level in the bottle reaches the beginning of the thread.
3
Carefully, without spilling water, push the cation exchanger bottle
over the inlet filter holder [D] into the bottle holder [B].
4
Screw the cation exchanger bottle into the bottle holder.
Do not tighten the bottle too firmly, this could damage the
gasket.
Pre-rinse
set-up
If you have a pre-rinse set-up, proceed according to
“Install cation
exchanger bottle”
to install the second cation exchanger bottle.
A
B
C
D
E
Flow cell
Bottle holder
Cation exchanger bottle
Inlet filter holder
Empty bottle
A
B
C
E
D