![Swann AMI Codes-II Operator'S Manual Download Page 11](http://html1.mh-extra.com/html/swann/ami-codes-ii/ami-codes-ii_operators-manual_1411948011.webp)
AMI Codes-II
Product Description
A-96.250.571 / 050517
9
2.
Product Description
Application
Range
The AMI Codes-II analyzer is a complete monitoring system for the
automatic, continuous measurement and dosing control of chlorine
and other disinfectants based on the DPD colorimetric method
AWWA 4500 Cl-G and on EN ISO 7393-2.
It can be used for measuring disinfectants in:
Potable water
Swimming pools
Cooling water
Effluent
Seawater
It is also applicable for water containing additives like corrosion in-
hibitors, cyanuric acids and antiscalants.
For the determination of free chlorine, chlorine dioxide, bromine
and iodine the reagents:
Oxycon on-line DPD
Oxycon on-line Buffer
are needed.
The measurement of monochloramine or ozone requires an addi-
tional reagent:
Oxycon on-line KI; which is added to Oxycon on-line Buffer
Disinfectant
Measurement
Signal
Outputs
Two signal outputs programmable for measured values (freely
scaleable, linear or bilinear) or as continuous control output (control
parameters programmable).
Current loop:
0/4–20 mA
Maximal burden:
510
Third signal output available as an option. The third signal output
can be operated as a current source or as a current sink (selectable
via switch).
Disinfectant
Measuring range
Accuracy
Ozone
0.000–1.000 ppm
± 0.005 ppm
HOCl / free chlorine /
Monochloramine
0.00–1.00 ppm
1.00–3.00 ppm
3.00–5.00 ppm
± 0.01 ppm
± 0.06 ppm
± 0.20 ppm
Chlorine dioxide /
Iodine / Bromine
0.00–2.00 ppm
2.00–6.00 ppm
± 0.02 ppm
± 0.12 ppm