AMI Codes-II CC
Maintenance
6.
Maintenance
6.1.
Maintenance Schedule
WARNING
Stop operation before maintenance.
Stop sample flow.
Shut off power of the instrument.
If option pH is installed
Daily (dirty water)
up to every
2 weeks (clean
water)
Check sample supply for dirt.
Clean all filters and strainers, if necessary.
Clean AMI Codes protection filter, if necessary.
Check sample flow (see also
)
Every 4–6 weeks
Clean reagent containers and prepare new reagents.
Let instrument run for 1 h. Make 3 manual measure-
ments. Compare average value to displayed value. If nec-
essary, perform process calibration.
Monthly
Recommendation: Check photometer with verification kit
Yearly
Exchange reagent pump tubes.
By occurrence:
E020, FOME dirty:
Cleaning the Photometer, p. 55
E022, Reagent empty:
Refill or replace Reagents, p. 46
E065, Reagents low:
Refill or replace Reagents, p. 46
Weekly
Perform a process calibration
,
see
Every 2 months
Perform a standard calibration
,
see
.