![Swan Analytical Instruments AMI Turbiwell Operator'S Manual Download Page 86](http://html1.mh-extra.com/html/swan-analytical-instruments/ami-turbiwell/ami-turbiwell_operators-manual_1411373086.webp)
A-96.250.511 / 091120
AMI Turbiwell
Program List and Explanations
86
9.
Program List and Explanations
1 Messages
1.1 Pending Errors
Provides the list of active errors with their status (active, acknowl
-
edged). If an active error is acknowledged, the alarm relay is active
again. Cleared errors are moved to the Message list.
1.2 Message List
Shows the error history: Error code, date and time of issue, and sta
-
tus (active, acknowledged, cleared).
64 errors are memorized. Then the oldest error is cleared to save the
newest error (circular buffer).
1.3 Maintenance List
Provides the list of necessary maintenance. Cleared maintenance
messages are moved to the Message list.
2 Diagnostics
In diagnostics mode, the values can only be viewed, not modified.
2.1 Identification
2.1.1
Designation
: Designation of the instrument.
2.1.2
Version
: Firmware of the instrument (e.g. V6.23-09/19).
2.1.3
Version
TURBI
, e.g. 1.35.
2.1.4
Factory test:
Test date of the instrument, motherboard and front-end
QC test.
2.1.5
Operating time:
years, days, hours, minutes, seconds.
2.2 Sensors
2.2.1
Turbidity:
Turbidity
: in FNU/NTU.
Raw value
: Turbidity in counts.
Quotient
: Division of detector signal by reference signal.
2.2.2
Miscellaneous
2.2.2.1
Case Temp.
: Current temperature in °C of electronics housing.
2.2.3
History
2.2.3.1
Cal. History:
Only for diagnostic purposes. Review the values of the
last calibrations. Max. 64 data records are memorized.