A-96.250.511 / 091120
AMI Turbiwell
Maintenance
56
6.2. Cleaning the Measuring Chamber
CAUTION
Wrong measured values due to dirty optical components.
Touching the optical components in the cover of the measuring
cell may result in wrong measured values.
Cleaning and re-calibration at SWAN is necessary.
Never touch the optical components during maintenance work.
Cleaning
To clean the measuring chamber proceed as follows:
1
Stop sample flow.
2
Release the two quick fastener screws [C].
The cover automatically slides up.
3
Move the cover completely backwards.
4
If the measuring chamber is strongly polluted, remove algae etc.
from the barrier and measuring chamber walls with a soft brush.
A
B
C
D
E
Drain valve
Measuring chamber
Quick fastener screw
Cover
Panel
A B C
D
E